19 CSR 20-10.160
Special Requirements for Mobile Medical Radiographic Installations
PURPOSE: This rule establishes special
requirements for mobile medical radiographic installations.
(1) All mobile equipment shall be provided
with cones or frames so that the minimum
target-to-skin distance is at least eighteen
inches (18").
(2) The exposure control switch shall be so
arranged that the operator can stand at least
six feet (6') from the patient and outside the
primary beam for all exposures.
(3) A device shall be provided to terminate
the exposure after a preset time.
AUTHORITY: section 192.420, RSMo 1986.*
This rule was previously filed as 13 CSR 5090.160. Original rule filed Nov. 9, 1964,
effective Dec. 9, 1964.
*Original authority: 192.420, RSMo 1963.