19 CSR 20-10.170
Special Requirements for Photofluorographic Installations
PURPOSE: This rule establishes special
requirements for photofluorographic installations.
(1) A collimator shall restrict the useful beam
to the area of the fluorographic screen.
(2) The exposure switch shall be so arranged
that it cannot be conveniently operated outside a shielded area. Exposure switches for
spot film devices used in conjunction with
fluoroscopic tables are exempted from this
requirement.
(3) A device shall be provided to terminate
the exposure after a preset time or exposure.
(4) Output of the photofluorographic installation shall not exceed one (1) roentgen per
exposure at the panel.
AUTHORITY: section 192.420, RSMo 1986.*
This rule was previously filed as 13 CSR 5090.170. Original rule filed Nov. 9, 1964,
effective Dec. 9, 1964.
*Original authority: 192.420, RSMo 1963.