19 CSR 30-11.070
Mammography Equipment Requirements
PURPOSE: This rule establishes requirements for mammography equipment, including radiation machines and other components
of mammography systems.
Editor's Note: The secretary of state has
determined that the publication of this rule in
its entirety would be unduly cumbersome or
expensive. The entire text of the material referenced has been filed with the secretary of
state. This material may be found at the
Office of the Secretary of State or at the headquarters of the agency and is available to any
interested person at a cost established by
state law.
(1) The equipment used for mammography
shall be specifically designed for mammography and shall meet the following requirements:
(A) Equipment shall meet the Food and
Drug Administration (FDA) performance
standards for diagnostic X-ray systems and
their major components in 21 CFR 1020.30
and FDA standards for radiographic equipment in 21 CFR 1020.31 upon installation
and while in use;
(B) The image receptor systems and all
their individual components shall be designed
specifically for mammography and shall
function properly;
(C) The equipment shall be limited to providing kilovoltage (kV)-target-filter combinations appropriate to image receptors meeting
the requirements of subsection (1)(B);
(D) The half value layer (HVL) in millimeters of aluminum of the useful beam for
systems operating at X-ray tube potentials of
less than thirty-five kilovoltage potential (35
kVp) shall be equal to or greater than the
product of the tube potential in kilovolts multiplied by one-hundredth (0.01) and shall not
exceed this calculated product by more than
plus one-tenth millimeter (+0.1 mm) of aluminum;
(E) Devices parallel to the imaging plane
shall be available to adequately immobilize
and compress the breast;
(F) Film-screen units shall have the capability for using anti-scatter grids;
(G) Film-screen units shall have the capability of automatic exposure control;
(H) The equipment shall have a control
panel that includes a device, usually a milliammeter, or a means for an audible signal to
give positive indication of the production of
X rays whenever the X-ray tube is energized.
The control panel shall include appropriate
indicators—labeled control settings or meters
that show the physical factors, such as kVp,
milliamperes (mA), milliampere-seconds
(mAs), exposure time or whether timing is
automatic—when used for exposure;
(I) The developer temperature of the photographic processor shall be maintained as
specified by the film manufacturer for the
particular film-developer-processor-development time combination; and
(J) The focal spot size and source-to-image
receptor distances shall be limited to those
appropriate for mammography.
AUTHORITY: sections 192.762 and 192.766,
RSMo Supp. 1992.* This rule was previously
filed as 19 CSR 20-11.070. Emergency rule
filed Oct. 5, 1992, effective Oct. 15, 1992,
expired Feb. 12, 1993. Emergency rule filed
Feb. 3, 1993, effective Feb. 13, 1993, expired
June 12, 1993. Original rule filed Nov. 4,
1992, effective June 7, 1993. Amended: Filed
April 28, 1993, effective Nov. 8, 1993.
Changed to 19 CSR 30-11.070 July 30, 1998.
*Original authority 1992.