19 CSR 30-11.080
Quality Assurance
PURPOSE: This rule establishes requirements for calibration of mammography equipment and monitoring of the mammography
system.
(1) All variable parameters of the equipment
shall be calibrated as follows:
(A) When the equipment is first installed;
(B) After any major changes or replacement of parts;
(C) At least annually while in use; and
(D) When quality assurance tests indicate
that calibration is needed.
(2) The following parameters of the system
shall be monitored as follows:
(A) Processor performance shall be monitored on a daily basis to include solution temperatures, sensitometric-densitometric evaluation through the use of control charts which
shall include speed step or mid-density step
with a control limit of plus or minus fifteenhundredths optical density (±0.15 OD), contrast index or density difference with a control limit of plus or minus fifteen-hundredths
optical density (±0.15 OD) and base plus fog
with a control limit of plus three-hundredths
optical density (+.03 OD) of the operating
level;
(B) Processor chemical replenishment
rates shall be monitored at least quarterly;
(C) Half-value layer shall be determined at
least annually with the compression device
located in the primary beam, halfway
between the image receptor assembly and the
X-ray tube;
(D) Milliampere-seconds (mAs) linearity
shall be determined at least annually based on
the average ratios of exposure to the indicated milliampere-seconds product (mR/mAs)
obtained at any two (2) consecutive tube current settings, which shall not differ by more
than ten-hundredths (0.10) times their sum as
measured with a minimum of four (4) consecutive exposures or a maximum of ten (10)
consecutive exposures at commonly used settings;
(E) Manual and automatic exposure control (AEC) reproducibility shall be determined at least annually based on the coefficient of variation of exposures, which shall
not be greater than plus or minus five percent
(±5%) as measured with a minimum of four
(4) consecutive exposures or a maximum of
ten (10) consecutive exposures at a commonly used setting;
(F) Kilovoltage potential (kVp) accuracy
shall be determined at least annually. The
actual kVp shall be maintained within plus or
minus five percent (±5%) of the indicated
kVp;
(G) kVp reproducibility shall be determined at least annually based on the coefficient of variation of kVp values, which shall
not be greater than plus or minus five percent
(±5%) as measured with a minimum of four
(4) consecutive exposures or a maximum of
ten (10) consecutive exposures at commonly
used kVp settings;
(H) AEC thickness response shall be determined at least annually based on film density, which shall be maintained within plus or
minus three-tenths optical density (±0.3 OD)
of the average OD over the kVp range used
for phantom thicknesses of two, four and six
centimeters (2, 4 and 6 cm). If the OD cannot be maintained to within plus or minus
three-tenths (‰0.3) of the average of each
clinically used setting, a technique chart shall
be developed that alters kVp and density control settings as a function of breast thicknesses and densities to produce optical densities
within this range under phototimed conditions;
(I) Compression device response shall be
monitored at least annually;
(J) Adequacy of film storage shall be monitored at least annually, both before use and
after exposure, if processing does not occur
immediately;
(K) Availability and use of technique charts
that shall include an indication of the kV-target-filter combination to be used with each
image receptor shall be monitored at least
annually;
(L) The use of kV-target-filter combination
appropriate to the image receptor shall be
monitored daily before patient irradiation;
(M) Darkroom integrity, for light-tight
conditions and use of proper safelight, shall
be monitored at least semiannually and whenever bulbs or filters are changed or when fog
is suspected; darkroom fog levels shall have a
variance of not greater than five-hundredths
(0.05) density units between the fogged and
unfogged film based on a two (2)-minute
exposure to the darkroom safelight;
(N) Image quality, using one (1) of the following mammography phantoms; the RMI
152, the RMI 156, the CIRS, the Nuclear
Associates phantom or other equivalent phantom acceptable to the Department of Health,
which simulates the composition of the breast
and provides indicators of disease conditions
allowing objective analysis of clinical image
quality, shall be monitored at least monthly
and each time the mammographic equipment
is moved or altered in any major way including the replacement of parts; a phantom
image shall be made and evaluated for mobile
units and vans at each location prior to the
performance of mammography;
(O) Average glandular dose shall be determined at least annually based on measurements of estimated skin entrance exposure;
the average glandular dose shall be calculated
from those used for imaging a four and fivetenths centimeters (4.5 cm) compressed
breast in the cranial caudal view and the measurement point shall be four and five-tenths
centimeters (4.5 cm) from the surface of the
grid or cassette as appropriate; and
(P) Film-screen contact shall be monitored
at least semiannually and when new or
repaired cassettes/screens are placed in service, using a contact test tool specifically
designed for mammography; cassettes/
screens shall be cleaned at least weekly; for
xeroradiography, the photoreceptor plates
shall be maintained in accordance with the
manufacturer’s specifications.
(3) Evaluation of monitoring results—
(A) Standards of image quality giving
acceptable ranges of value for each of the
parameters tested shall be established to aid
in the evaluation of monitoring results. The
standards of image quality related to dose
shall include a requirement that the average
glandular dose for one (1) cranial caudal
view of a four and five-tenths centimeters
(4.5 cm) compressed breast fifty percent
(50%) glandular/fifty percent (50%) adipose
composition shall not exceed one hundred
(100), three hundred (300) and four hundred
(400) millirads (mrads) for film/screen units
without grids, film/screen units with grids,
and xeroradiography units, respectively; and
(B) The monitoring results shall be compared routinely to the standards of image
quality established by the Department of
Health. If the results fall outside the acceptable range, the test shall be repeated. If the
results continue to be unacceptable, the
source of the problem shall be identified and
corrected before further examinations are
conducted.
(4) A program to analyze retakes shall be
established to detect and correct problems
affecting image quality or exposure. This
analysis shall be performed at least quarterly
or after two hunndred fifty (250) patients,
whichever comes first.
(5) All quality assurance test equipment shall
be in good operating order and calibrated
according to the manufacturer’s specifications.
AUTHORITY: sections 192.762 and 192.766,
RSMo. Supp. 1992.* This rule was previously filed as 19 CSR 20-11.080. Emergency rule
filed Oct. 5, 1992, effective Oct. 15, 1992,
expired Feb. 12, 1993. Emergency rule filed
Feb. 3, 1993, effective Feb. 13, 1993, expired
June 12, 1993. Original rule filed Nov. 4,
1992, effective June 7, 1993. Changed to 19
CSR 30-11.080 July 30, 1998.
*Original authority 1992.