19 CSR 30-11.090
Maintenance of Mammography Records
PURPOSE: This rule establishes requirements for maintenance of quality assurance
records by the supplier.
(1) Records of all quality assurance tests for
mammography X-ray equipment shall include
the following:
(A) The initial mammographic equipment
performance evaluation and subsequent evaluations and testing;
(B) Verification that the mammographic Xray equipment is in safe operating order; and
(C) Quality assurance test results.
(2) Records or written logs of maintenance
and any repairs of mammographic X-ray
equipment shall be kept for at least three (3)
years.
(3) Film processor control charts used to regulate proper processor function shall be kept
for at least two (2) years.
(4) Film processor maintenance logs shall
include all of the following records and these
records shall be kept for at least two (2)
years: preventive maintenance; corrective
maintenance; and cleaning and replacement
of chemicals.
(5) Each film processor record entry shall be
dated and signed or initialed by the individual who performed the quality assurance test.
(6) Records of calibration, monitoring
results, the radiation physicist’s on-site consultation and the physician consultant’s annual review shall be documented in the supplier’s records and shall be readily available for
review by agents of the Department of
Health.
AUTHORITY: sections 192.762 and 192.766,
RSMo Supp. 1992.* This rule was previously
filed as 19 CSR 20-11.090. Emergency rule
filed Oct. 5, 1992, effective Oct. 15, 1992,
expired Feb. 12, 1993. Emergency rule filed
Feb. 3, 1993, effective Feb. 13, 1993, expired
June 12, 1993. Original rule filed Nov. 4,
1992, effective June 7, 1993. Changed to 19
CSR 30-11.090 July 30, 1998.