15 MAC Pt. 21, R. 1.15.6
Therapeutic Radiation Machines of Less Than 500 kV
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.15.6
Therapeutic Radiation Machines of Less Than 500 kV.
1.
Leakage Radiation. When the x-ray tube is operated at its maximum rated tube
current for the maximum kV, the leakage kerma rate shall not exceed the value
specified at the distance specified for that classification of therapeutic radiation
machine:
a.
5-50 kV Systems. The leakage air kerma rate measured at any position 5
centimeters from the tube housing assembly shall not exceed 100 mrad (1
mGy) in any one hour.
b.
>50 and <500 kV Systems. The leakage air kerma rate measured at a
distance of 1 meter from the source in any direction shall not exceed 1 rad
(1 cGy) in any one hour. This air kerma rate measurement may be averaged
over areas no larger than 100 square centimeters. In addition, the air kerma
rate at a distance of 5 centimeters from the surface of the tube housing
assembly shall not exceed 30 rad (3O cGy) per hour.
2.
Permanent Beam-Limiting Devices. Permanent diaphragms or cones used for
limiting the useful beam shall provide at least the same degree of attenuation as
required for the tube housing assembly.
3.
Adjustable or Removable Beam-Limiting Devices.
a.
All adjustable or removable beam-limiting devices, diaphragms, cones or
blocks shall not transmit more than 5 percent of the useful beam for the
most penetrating beam used.
b.
When adjustable beam-limiting devices are used, the position and shape of
the radiation field shall be indicated by a light beam.
4.
Filter System. The filter system shall be so designed that:
a.
Filters cannot be accidentally displaced at any possible tube orientation;
b.
For equipment installed after the effective date of these regulations, an
interlock system prevents irradiation if the proper filter is not in place;
c.
The air kerma rate escaping from the filter slot shall not exceed 1 rad (1
cGy) per hour under any operating conditions; and
d.
Each filter shall be marked as to its material of construction and its
thickness.
5.
Tube Immobilization.
a.
The x-ray tube shall be so mounted that it cannot accidentally turn or slide
with respect to the housing aperture; and
b.
The tube housing assembly shall be capable of being immobilized for
stationary portal treatments.
6.
Source Marking. The tube housing assembly shall be so marked that it is possible
to determine the location of the source to within 5 millimeters, and such marking
shall be readily accessible for use during calibration procedures.
7.
Beam Block. Contact therapy tube housing assemblies shall have a removable
shield of material, equivalent in attenuation to 0.5 millimeters of lead at 100 kV,
which can be positioned over the entire useful beam exit port during periods when
the beam is not in use.
8.
Timer. A suitable irradiation control device shall be provided to terminate the
irradiation after a preset time interval.
a.
A timer which has a display shall be provided at the treatment control panel.
The timer shall have a preset time selector and an elapsed time indicator;
b.
The timer shall be a cumulative timer which activates with an indication of
"BEAM-ON" and retains its reading after irradiation is interrupted or
terminated. After irradiation is terminated and before irradiation can be
reinitiated, it shall be necessary to reset the elapsed time indicator;
c.
The timer shall terminate irradiation when a preselected time has elapsed,
if any dose monitoring system present has not previously terminated
irradiation;
d.
The timer shall permit accurate presetting and determination of exposure
times as short as 1 second;
e.
The timer shall not permit an exposure if set at zero;
f.
The timer shall not activate until the shutter is opened when irradiation is
controlled by a shutter mechanism unless calibration includes a timer error
correction to compensate for mechanical lag; and
g.
The timer shall be accurate to within 1 percent of the selected value or 1
second, whichever is greater.
9.
Control Panel Functions. The control panel, in addition to the displays required
by other provisions in 1.15.6 shall have:
a.
An indication of whether electrical power is available at the control panel
and if activation of the x-ray tube is possible;
b.
An indication of whether x-rays are being produced;
c.
Means for indicating x-ray tube potential and current;
d.
The means for terminating an exposure at any time;
e.
A locking device which will prevent unauthorized use of the therapeutic
radiation machine; and
f.
For therapeutic radiation machines manufactured after the effective date of
these regulations, a positive display of specific filter(s) in the beam.
10.
Multiple Tubes. When a control panel may energize more than one x-ray tube:
a.
It shall be possible to activate only one x-ray tube at any time;
b.
There shall be an indication at the control panel identifying which x-ray
tube is activated; and
c.
There shall be an indication at the tube housing assembly when that tube is
energized.
11.
Source-to-Skin Distance (SSD). There shall be a means of determining the central
axis SSD to within 1 centimeter and of reproducing this measurement to within 2
millimeters thereafter.
12.
Shutters. Unless it is possible to bring the x-ray output to the prescribed exposure
parameters within 5 seconds after the x-ray "ON" switch is energized, the beam
shall be attenuated by a shutter having a lead equivalency not less than that of the
tube housing assembly. In addition, after the unit is at operating parameters, the
shutter shall be controlled electrically by the operator from the control panel. An
indication of shutter position shall appear at the control panel.
13.
Low Filtration X-ray Tubes. Each therapeutic radiation machine equipped with
a beryllium or other low-filtration window shall be clearly labeled as such upon the
tube housing assembly and shall be provided with a permanent warning device on
the control panel that is activated when no additional filtration is present to indicate
that the dose rate is very high.
14.
Facility Design Requirements for Therapeutic Radiation Machines Capable of
Operating in the Range 50 kV to 500 kV. In addition to shielding adequate to
meet requirements of 1.15.9, the treatment room shall meet the following design
requirements:
a.
Aural Communication. Provision shall be made for continuous two-way
aural communication between the patient and the operator at the control
panel.
b.
Viewing Systems. Provision shall be made to permit continuous
observation of the patient during irradiation and shall be so located that the
operator can observe the patient from the control panel. The therapeutic
radiation machine shall not be used for patient irradiation unless at least one
viewing system is operational.
15.
Additional Requirements. Treatment rooms which contain a therapeutic radiation
machine capable of operating above 150 kV shall meet the following additional
requirements:
a.
All protective barriers shall be fixed except for entrance doors or beam
interceptors;
b.
The control panel shall be located outside the treatment room or in a totally
enclosed booth, which has a ceiling, inside the room;
c.
Interlocks shall be provided such that all entrance doors, including doors to
any interior booths, shall be closed before treatment can be initiated or
continued. If the radiation beam is interrupted by any door opening, it shall
not be possible to restore the machine to operation without closing the door
and reinitiating irradiation by manual action at the control panel; and
d.
When any door referred to in 1.15.6(15)(c) is opened while the x-ray tube
is activated, the air kerma rate at a distance of 1 meter from the source shall
be reduced to less than 100 mrad (1 mGy) per hour.
16.
Full Calibration Measurements.
a.
Full calibration of a therapeutic radiation machine subject to 1.15.6 shall be
performed by, or under the direct supervision of, a Radiation Therapy
Physicist:
i.
Before the first medical use following installation or reinstallation
of the therapeutic radiation machine;
ii.
At intervals not exceeding 1 year; and
iii.
Before medical use under the following conditions:
i.
Whenever quality assurance check measurements indicate
that the radiation output differs by more than 5 percent from
the value obtained at the last full calibration and the
difference cannot be easily discerned; and
ii.
Following any component replacement, major repair, or
modification of components that could significantly affect
the characteristics of the radiation beam.
b.
To satisfy the requirement of 1.15.6(16)(a), full calibration shall include all
measurements recommended for annual calibration by NCRP Report 69,
"Dosimetry of X-Ray and Gamma Ray Beams for Radiation Therapy in the
Energy Range 10 keV to 50 MeV" (1981).
c.
A registrant shall maintain a record of each calibration for the duration of
the registration. The record shall include the date of the calibration, the
manufacturer's name, model number, and serial number for both the
therapeutic radiation machine and the x-ray tube, the model numbers and
serial numbers of the instruments used to calibrate the therapeutic radiation
machine, and the signature of the Radiation Therapy Physicist responsible
for performing the calibration.
17.
Periodic Quality Assurance Checks.
a.
Periodic quality assurance checks shall be performed on therapeutic
radiation machines subject to 1.15.6, which are capable of operation at
greater than 150 kV.
b.
To satisfy the requirement of 1.15.6(17)(a), quality assurance checks shall
meet the following requirements:
i.
The registrant shall perform quality assurance checks in accordance
with written procedures established by the Radiation Therapy
Physicist; and
ii.
The quality assurance check procedures shall specify the frequency
at which tests or measurements are to be performed. The quality
assurance check procedures shall specify that the quality assurance
check shall be performed during the calibration specified
1.15.6(16)(a). The acceptable tolerance for each parameter
measured in the quality assurance check, when compared to the
value for that parameter determined in the calibration specified in
1.15.6(16)(a) shall be stated.
c.
The cause for a parameter exceeding a tolerance set by the Radiation
Therapy Physicist shall be investigated and corrected before the system is
used for patient irradiation.
d.
Whenever a quality assurance check indicates a significant change in the
operating characteristics of a system, as specified in the Radiation Therapy
Physicist's quality assurance check procedures, the system shall be
recalibrated as required in 1.15.6(16)(a).
e.
The registrant shall use the dosimetry system described in 1.15.4(3) to make
the quality assurance check required in 1.15.6(17)(b).
f.
The registrant shall have the Radiation Therapy Physicist review and sign
the results of each radiation output quality assurance check at intervals not
to exceed 1 month. The Radiation Therapy Physicist shall promptly notify
the registrant in writing of the results of each radiation output quality
assurance check. The registrant shall keep a copy of each written
notification for 3 years.
g.
Therapeutic radiation machines subject to 1.15.6 shall have safety quality
assurance checks of each external beam radiation therapy facility performed
at intervals not to exceed 1 month.
h.
To satisfy the requirement of 1.15.6(17)(g), safety quality assurance checks
shall ensure proper operation of:
i.
Electrical interlocks at each external beam radiation therapy room
entrance;
ii.
Proper operation of the "BEAM-ON" and termination switches;
iii.
Beam condition indicator lights on the access door(s), control
console, and in the radiation therapy room;
iv.
Viewing systems; and
v.
Electrically operated treatment room doors from inside and outside
the treatment room.
i.
The registrant shall promptly repair any system identified in 1.15.6(17)(h)
that is not operating properly.
j.
The registrant shall maintain a record of each quality assurance check
required by 1.15.6(17)(a) and 1.15.6(17)(g) for 3 years. The record shall
include the date of the quality assurance check, the manufacturer's name,
model number, and serial number for the therapeutic radiation machine, the
manufacturer's name, model number and serial number of the instrument(s)
used to measure the radiation output of the therapeutic radiation machine,
and the signature of the individual who performed the periodic quality
assurance check.
18.
Operating Procedures.
a.
The therapeutic radiation machine shall not be used for irradiation of
patients unless the requirements of 1.15.6(16) and 1.15.6(17) have been
met;
b.
Therapeutic radiation machines shall not be left unattended unless it is
secured pursuant to 1.15.6(9)(e);
c.
When a patient must be held in position for radiation therapy, mechanical
supporting or restraining devices shall be used;
d.
The tube housing assembly shall not be held by an individual during
operation unless the assembly is designed to require such holding and the
peak tube potential of the system does not exceed 50 kV. In such cases, the
holder shall wear protective gloves and apron of not less than 0.5
millimeters lead equivalency at 100 kV;
e.
A copy of the current operating and emergency procedures shall be
maintained at the therapeutic radiation machine control console; and
f.
No individual other than the patient shall be in the treatment room during
exposures from therapeutic radiation machines operating above 150 kV. At
energies less than or equal to 150 kV, any individual, other than the patient,
in the treatment room shall be protected by a barrier sufficient to meet the
requirements of 1.4.6 of these regulations.
19.
Possession of Survey Instrument(s). The registrant authorized to use a
therapeutic radiation machine in accordance with 1.15.6 shall possess appropriately
calibrated portable monitoring equipment. As a minimum, such equipment shall
include a portable radiation measurement survey instrument capable of measuring
dose rates over the range 1 mrem (10 Sv) per hour to 1000 mrem (10 mSv) per
hour. The survey instrument(s) shall be operable and calibrated in accordance with
1.15.8.