15 MAC Pt. 21, R. 1.15.7
Therapeutic Radiation Machines - Photon Therapy Systems (500 kV and
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.15.7
Therapeutic Radiation Machines - Photon Therapy Systems (500 kV and
Above) and Electron Therapy Systems (500 keV and Above).
1.
Possession of Survey Instrument(s). A registrant authorized to use a therapeutic
radiation machine in accordance with 1.15.7 shall possess appropriately calibrated
portable monitoring equipment. As a minimum, such equipment shall include a
portable radiation measurement survey instrument capable of measuring dose rates
over the range 1 mrem (10 Sv) per hour to 1000 mrem (10 mSv) per hour. The
survey instrument(s) shall be operable and calibrated in accordance with 1.15.8.
2.
Leakage Radiation Outside the Maximum Useful Beam in Photon and
Electron Modes.
a.
The absorbed dose rate due to leakage radiation (excluding neutrons) at any
point outside the maximum sized useful beam, but within a circular plane
of radius 2 meters which is perpendicular to and centered on the central axis
of the useful beam at the nominal treatment distance (i.e., patient plane),
shall not exceed a maximum of 0.2 percent and an average of 0.1 percent of
the absorbed dose rate on the central axis of the beam at the nominal
treatment distance. Measurements shall be averaged over an area up to but
not exceeding 100 square centimeters at the positions specified;
b.
Except for the area defined in 1.15.7(2)(a), the absorbed dose rate in tissue
(excluding that from neutrons) at I meter from the electron path between the
source and the target or electron window shall not exceed 0.5 percent of the
absorbed dose rate in tissue on the central axis of the beam at the nominal
treatment distance. Measurements shall be averaged over an area up to but
not exceeding 100 square centimeters at the positions specified;
c.
The neutron absorbed dose rate outside the useful beam shall be kept as low
as practicable. Measurements of the portion of the leakage radiation dose
contributed by neutrons shall be averaged over an area up to but not
exceeding 800 square centimeters; and
d.
For each therapeutic radiation machine, the registrant shall determine, or
obtain from the manufacturer, the leakage radiation existing at the positions
specified in 1.15.7(2)(a) through 1.15.7(2)(c) for the specified operating
conditions. Records on leakage radiation measurements shall be maintained
at the installation for inspection by the Agency.
3.
Leakage Radiation Through Beam-Limiting Devices.
a.
Photon Radiation. All adjustable or interchangeable beam-limiting devices
shall attenuate the useful beam such that:
i.
At the nominal treatment distance, the maximum absorbed dose
anywhere in the area shielded by the beam-limiting device(s) shall
not exceed 2 percent of the maximum absorbed dose on the central
axis of the useful beam measured in a 10 centimeters by 10
centimeters radiation field.
ii.
For fields of any size in which the maximum area shielded by the
beam-limiting devices exceeds 500 square centimeters, the product
of the average absorbed dose due to leakage radiation through the
beam-limiting devices and the maximum area protectable by the
beam-limiting devices shall not exceed one tenth (0.1) of the product
of the maximum absorbed dose on the central axis of the useful
beam and the area of the useful beam for a radiation field of 10
centimeters by 10 centimeters. All values of absorbed dose and area
are referred to the nominal treatment distance.
b.
Electron Radiation. All adjustable or interchangeable electron applicators
shall attenuate the radiation, including but not limited to photon radiation
generated by electrons incident on the beam-limiting device and electron
applicator and other parts of the radiation head, such that the following
limits apply:
i.
The absorbed dose in a plane perpendicular to the central axis of the
useful beam at the nominal treatment distance shall not exceed:
i.
An average of 2 percent of the absorbed dose on the central
axis of the useful beam at the nominal treatment distance.
This limit shall apply in the area between a line 4 centimeters
outside the periphery of the geometrical radiation field and
the border of the maximum area protectable by the electron
applicator; and
ii.
A maximum of 10 percent of the absorbed dose on the
central axis of the useful beam at the nominal treatment
distance. This limit shall apply in the area between a line 2
centimeters outside the periphery of the geometrical
radiation field and the border of the maximum area
protectable by the electron applicator.
ii.
For fields of any size in which the maximum area shielded
by the electron applicator exceeds 1000 square centimeters,
the product of the average absorbed dose due to leakage
radiation through the electron applicators and the maximum
area protectable by the electron applicators shall not exceed
two tenths (0.2) of the product of the maximum absorbed
dose on the central axis of the useful beam and the area of
the useful beam for a radiation field of 10 centimeters by 10
centimeters. All values of absorbed dose and area are
referred to the nominal treatment distance.
c.
Measurement of Leakage Radiation.
i.
Photon Radiation. Measurements of leakage radiation through the
beam-limiting devices shall be made with the beam-limiting devices
closed and any residual aperture blocked by at least two (2) tenth
value layers of suitable absorbing material. In the case of
overlapping beam-limiting devices, the leakage radiation through
each set shall be measured independently, and the leakage radiation
from each set shall not exceed a maximum of 2 percent anywhere in
the area protectable by that beam-limiting device.
ii.
Electron Radiation. Measurements of leakage radiation through
the electron applicators shall be made with the electron beam
directed into the air and using a radiation detector of area up to but
not exceeding 1 square centimeter suitably protected against
radiation which has been scattered from material beyond the
radiation detector. Measurements shall be made using 1 centimeter
of tissue equivalent build up material.
d.
When adjustable beam-limiting devices are used, the position and shape of
the radiation field shall be indicated by a light field.
4.
Filters/Wedges.
a.
Each filter and/or wedge which is removable from the system shall be
clearly marked with an identification number. For removable wedge filters,
the nominal wedge angle shall appear on the wedge or wedge tray (if
permanently mounted to the tray). If the wedge or wedge tray is damaged,
the wedge transmission factor shall be redetermined;
b.
If the absorbed dose rate information required by 1.15.7(9) relates
exclusively to operation with a field flattening or beam scattering filter in
place, such filter shall be removable only by the use of tools;
c.
For equipment manufactured after the effective date of these regulations
which utilize a system of wedge filters, interchangeable field flattening
filters, or interchangeable beam scattering foils:
i.
Irradiation shall not be possible until a selection of a filter or a
positive selection to use "no filter" has been made at the treatment
control panel, either manually or automatically;
ii.
An interlock system shall be provided to prevent irradiation if the
filter selected is not in the correct position;
iii.
A display shall be provided at the treatment control panel showing
the wedge filter(s), interchangeable field flattening filter(s), and/or
interchangeable beam scattering foil(s) in use; and
iv.
An interlock shall be provided to prevent irradiation if any filter
and/or beam scattering foil selection operation carried out in the
treatment room does not agree with the filter and/or beam scattering
foil selection operation carried out at the treatment control panel.
5.
X-Ray Stray Radiation in the Useful Electron Beam. For equipment
manufactured after the effective date of these regulations, the registrant shall
determine during acceptance testing, or obtain from the manufacturer, data
sufficient to ensure that x-ray stray radiation in the useful electron beam, absorbed
dose at the surface during x-ray irradiation and stray neutron radiation in the useful
x-ray beam are in compliance with International Electrotechnical Commission
(IEC) Document 601-2-1 (most current revision).
6.
Beam Monitors. All therapeutic radiation machines subject to 1.15.7 shall be
provided with beam monitoring devices. The sensors for this device shall be fixed
in the useful beam during treatment, (or interlocked) to indicate the air kerma rate
or dose rate.
a.
Equipment manufactured after the effective date of these regulations shall
be provided with at least two independently powered integrating dose
meters. Alternatively, a common power supply may be used if the
production of radiation is terminated upon failure of any common element.
b.
Equipment manufactured on or before the effective date of these regulations
shall be provided with at least one radiation detector. This detector shall be
incorporated into a primary beam monitoring system;
c.
The detector and the system into which that detector is incorporated shall
meet the following requirements:
i.
Each detector shall be removable only with tools and, if movable,
shall be interlocked to prevent incorrect positioning;
ii.
Each detector shall form part of a beam monitoring system from
whose readings in dose monitor units the absorbed dose at a
reference point in the treatment volume can be calculated;
iii.
Each beam monitoring system shall be capable of independently
monitoring, interrupting, and terminating irradiation; and
iv.
For equipment manufactured after the effective date of these
regulations, the design of the beam monitoring systems shall ensure
that the:
i.
Malfunctioning of one system shall not affect the correct
functioning of the secondary system; and
ii.
Failure of any element common to both systems which could
affect the correct function of both systems shall terminate
irradiation or prevent the initiation of radiation.
v.
Each beam monitoring system shall have a legible display at the
treatment control panel. For equipment manufactured after the
effective date of these regulations, each display shall:
i.
maintain a reading until intentionally reset;
ii.
have only one scale and no electrical or mechanical scale
multiplying factors;
iii.
utilize a design such that increasing dose is displayed by
increasing numbers; and
iv.
In the event of power failure, the beam monitoring
information required in 1.15.7(6)(c)(v)(iii) displayed at the
control panel at the time of failure shall be retrievable in at
least one system for a 20 minute period of time.
7.
Beam Symmetry.
a.
Bent-beam linear accelerators subject to 1.15.7 shall be provided with
auxiliary device(s) to monitor beam symmetry;
b.
The device(s) referenced in 1.15.7(7)(a) shall be able to detect field
asymmetry greater than 10 percent; and
c.
The device(s) referenced in 1.15.7(7)(a) shall be configured to terminate
irradiation if the specifications in 1.15.7(7)(b) cannot be maintained.
8.
Selection and Display of Dose Monitor Units.
a.
Irradiation shall not be possible until a selection of a number of dose
monitor units has been made at the treatment control panel;
b.
The preselected number of dose monitor units shall be displayed at the
treatment control panel until reset manually for the next irradiation;
c.
After termination of irradiation, it shall be necessary to reset the dosimeter
display before subsequent treatment can be initiated; and
d.
For equipment manufactured after the effective date of these regulations,
after termination of irradiation, it shall be necessary for the operator to reset
the preselected dose monitor units before irradiation can be initiated.
9.
Air Kerma Rate/Absorbed Dose Rate. For equipment manufactured after the
effective date of these regulations, a system shall be provided whose readings the
air kerma rate or absorbed dose rate at a reference point in the treatment volume
can be calculated. [The radiation detectors specified in 1.15.7(6) may form part of
this system.] In addition:
a.
The dose monitor unit dose rate shall be displayed at the treatment control
panel;
b.
If the equipment can deliver under any conditions, an air kerma rate or
absorbed dose rate at the nominal treatment distance more than twice the
maximum value specified by the manufacturer, a device shall be provided
which terminates irradiation when the air kerma rate or absorbed dose rate
exceeds a value twice the specified maximum. The dose rate at which the
irradiation will be terminated shall be a record maintained by the registrant;
and
c.
For equipment manufactured after the effective date of these regulations, if
the equipment can deliver under any conditions an air kerma rate or
absorbed dose rate at the nominal treatment distance more than ten (10)
times the maximum value specified by the manufacturer, a device shall be
provided to prevent the air kerma rate or absorbed dose rate anywhere in the
radiation field from exceeding twice the specified maximum value and to
terminate irradiation if the excess absorbed dose at the nominal treatment
distance exceeds 400 rad (4 Gy).
10.
Termination of Irradiation by the Beam Monitoring System or Systems
During Stationary Beam Radiation Therapy.
a.
Each primary system shall terminate irradiation when the preselected
number of dose monitor units has been detected by the system;
b.
If original design of the equipment included a secondary dose monitoring
system, that system shall be capable of terminating irradiation when not
more than 15 percent or 40 dose monitor units above the preselected number
of dose monitor units set at the control panel has been detected by the
secondary dose monitoring system; and
c.
For equipment manufactured after the effective date of these regulations, an
indicator on the control panel shall show which monitoring system has
terminated irradiation.
11.
Termination Switches. It shall be possible to terminate irradiation and equipment
movement or go from an interruption condition to termination condition at any time
from the operator's position at the treatment control panel.
12.
Interruption Switches. If a therapeutic radiation machine has an interrupt mode,
it shall be possible to interrupt irradiation and equipment movements at any time
from the treatment control panel. Following an interruption, it shall be possible to
restart irradiation by operator action without any reselection of operating
conditions. If any change is made of a preselected value during an interruption,
irradiation and equipment movements shall be automatically terminated.
13.
Timer. A suitable irradiation control device shall be provided to terminate the
irradiation after a preset time interval.
a.
A timer shall be provided which has a display at the treatment control panel.
The timer shall have a preset time selector and an elapsed time indicator;
b.
The timer shall be a cumulative timer which activates with an indication of
"BEAM-ON" and retains its reading after irradiation is interrupted or
terminated. After irradiation is terminated and before irradiation can be
reinitiated, it shall be necessary to reset the elapsed time indicator;
c.
For equipment manufactured after the effective date of these regulations,
after termination of irradiation and before irradiation can be reinitiated, it
shall be necessary for the operator to reset the preset time selector; and
d.
The timer shall terminate irradiation when a preselected time has elapsed,
if the dose monitoring systems have not previously terminated irradiation.
14.
Selection of Radiation Type. Equipment capable of both x-ray therapy and
electron therapy shall meet the following additional requirements:
a.
Irradiation shall not be possible until a selection of radiation type (x-rays or
electrons) has been made at the treatment control panel;
b.
The radiation type selected shall be displayed at the treatment control panel
before and during irradiation;
c.
An interlock system shall be provided to ensure that the equipment can
principally emit only the radiation type which has been selected;
d.
An interlock system shall be provided to prevent irradiation with x-rays
except to obtain a verification film, when electron applicators are fitted;
e.
An interlock system shall be provided to prevent irradiation with electrons
when accessories specific for x-ray therapy are fitted; and
f.
An interlock system shall be provided to prevent irradiation if any selected
operations carried out in the treatment room do not agree with the selected
operations carried out at the treatment control panel.
15.
Selection of Energy. Equipment capable of generating radiation beams of different
energies shall meet the following requirements:
a.
Irradiation shall not be possible until a selection of energy has been made
at the treatment control panel;
b.
The measured energy value selected shall be displayed (MV for photons
and MeV for electrons) at the treatment control panel before and during
irradiation; and
c.
Irradiation shall not be possible until the appropriate flattening filter or
scattering foil for the selected energy is in its proper location.
16.
Selection of Stationary Beam Radiation Therapy or Rotational Arc Radiation
Therapy. Therapeutic radiation machines capable of both stationary beam
radiation therapy and rotational arc radiation therapy shall meet the following
requirement:
a.
Irradiation shall not be possible until a selection of stationary beam radiation
therapy or rotational arc radiation therapy has been made at the treatment
control panel;
b.
The mode of operation shall be displayed at the treatment control panel;
c.
An interlock system shall be provided to ensure that the equipment can
operate only in the mode which has been selected;
d.
An interlock system shall be provided to prevent irradiation if any selected
parameter in the treatment room does not agree with the selected parameter
at the treatment control panel;
e.
Rotational arc radiation therapy shall be controlled to obtain the selected
relationships between incremental dose monitor units and incremental angle
of movement:
i.
For equipment manufactured after the effective date of these
regulations, an interlock system shall be provided to terminate
irradiation if the number of dose monitor units delivered in any 15
degrees of arc differs by more than 20 percent from the selected
value;
ii.
For equipment manufactured after the effective date of these
regulations, where gantry angle terminates the irradiation in
rotational arc radiation therapy, the dose monitor units shall differ
by less than 5 percent from the value calculated from the absorbed
dose per unit angle and total angle relationship;
iii.
For equipment manufactured after the effective date of these
regulations, an interlock shall be provided to prevent the gantry
moving more than 5 degrees beyond the selected angular limits
during rotational arc radiation therapy; and
iv.
For equipment manufactured after the effective date of these
regulations, an interlock shall be provided to require that a selection
of direction be made at the treatment control panel in all units which
are capable of both clockwise and counter-clockwise rotational arc
radiation therapy.
f.
Where the beam monitor system terminates the irradiation in rotational arc
radiation therapy, the termination of irradiation shall be as required by
1.15.7(10); and
g.
For equipment manufactured after the effective date of these regulations, an
interlock system shall be provided to terminate irradiation if movement of
the gantry:
i.
occurs during stationary beam radiation therapy; or
ii.
stops during rotational arc radiation therapy unless such stoppage is
a preplanned function.
17.
Facility Design Requirements for Therapeutic Radiation Machines Operating
above 500 kV. In addition to shielding adequate to meet requirements of 1.15.9,
the following design requirements are made:
a.
Protective Barriers. All protective barriers shall be fixed, except for
access doors to the treatment room or movable beam interceptors;
b.
Control Panel. In addition to other requirements specified in Subchapter
15, the control panel shall also:
i.
Be located outside the treatment room;
ii.
Provide an indication of whether electrical power is available at the
control panel and if activation of the radiation is possible;
iii.
Provide an indication of whether radiation is being produced; and
iv.
Include an access control (locking) device which will prevent
unauthorized use of the therapeutic radiation machine;
c.
Viewing Systems. Windows, mirrors, closed-circuit television or an
equivalent viewing system shall be provided to permit continuous
observation of the patient following positioning and during irradiation and
shall be so located that the operator may observe the patient from the
treatment control panel. The therapeutic radiation machine shall not be used
for patient irradiation unless at least one viewing system is operational;
d.
Aural Communications. Provision shall be made for continuous two-way
aural communication between the patient and the operator at the control
panel. The therapeutic radiation machine shall not be used for irradiation
of patients unless continuous two-way aural communication is possible;
e.
Room Entrances. Treatment room entrances shall be provided with
warning lights in a readily observable position near the outside of all access
doors, which will indicate when the useful beam is "ON" and when it is
"OFF";
f.
Entrance Interlocks. Interlocks shall be provided such that all entrance
doors must be closed before treatment can be initiated or continued. If the
radiation beam is interrupted by any door opening, it shall not be possible
to restore the machine to operation without closing the door and reinitiating
irradiation by manual action at the control panel;
g.
Beam Interceptor Interlocks. If the shielding material in any protective
barrier requires the presence of a beam interceptor to ensure compliance
with 1.4.14(1), and 1.4.14(2) of these regulations, interlocks shall be
provided to prevent the production of radiation, unless the beam interceptor
is in place, whenever the useful beam is directed at the designated barrier(s);
h.
Emergency Cutoff Switches. At least one (1) "scram button" or other
emergency power cutoff switch shall be located in the radiation therapy
room and shall terminate all equipment electrical power including radiation
and mechanical motion. This switch is in addition to the termination switch
required by 1.15.7(11). All emergency power cutoff switches shall include
a manual reset so that the therapeutic radiation machine cannot be restarted
from the unit's control console without resetting the emergency cutoff
switch;
i.
Safety Interlocks. All safety interlocks shall be designed so that any defect
or component failure in the safety interlock system prevents or terminates
operation of the therapeutic radiation machine; and
j.
Surveys for Residual Radiation. Surveys for residual activity, shall be
conducted on all therapeutic radiation machines capable of generating
photon and electron energies above 10 MV prior to machining, removing,
or working on therapeutic radiation machine components which may have
become activated due to photo-neutron production.
18.
Radiation Therapy Physicist Support.
a.
The services of a Radiation Therapy Physicist shall be utilized in facilities
having therapeutic radiation machines with energies of 500 kV and above.
The Radiation Therapy Physicist shall be responsible for:
i.
Full calibration(s) required by 1.15.7(20) and protection surveys
required by 1.15.4(1);
ii.
Supervision and review of dosimetry;
iii.
Beam data acquisition and storage for computerized dosimetry, and
supervision of its use;
iv.
Quality assurance, including quality assurance check review
required by 1.15.7(21)(e) of these regulations;
v.
Consultation with the authorized user in treatment planning, as
needed; and
vi.
Perform calculations/assessments regarding misadministrations.
b.
If the Radiation Therapy Physicist is not a full-time employee of the
registrant, the operating procedures required by 1.15.7(19) shall also
specifically address how the Radiation Therapy Physicist is to be contacted
for problems or emergencies, as well as the specific actions, if any, to be
taken until the Radiation Therapy Physicist can be contacted.
19.
Operating Procedures.
a.
No individual other than the patient shall be in the treatment room during
treatment of a patient or during any irradiation for testing or calibration
purposes;
b.
Therapeutic radiation machines shall not be made available for medical use
unless the requirements of 1.15.4(1), 1.15.7(20), and 1.15.7(21) have been
met;
c.
Therapeutic radiation machines, when not in operation, shall be secured to
prevent unauthorized use;
d.
If a patient must be held in position during treatment, mechanical supporting
or restraining devices shall be used; and
e.
A copy of the current operating and emergency procedures shall be
maintained at the therapeutic radiation machine control console.
20.
Full Calibration Measurements.
a.
Full calibration of a therapeutic radiation machine subject to 1.15.7 shall be
performed by, or under the direct supervision of, a Radiation Therapy
Physicist:
i.
Before the first medical use following installation or reinstallation
of the therapeutic radiation machine;
ii.
At intervals not exceeding 1 year; and
iii.
Before medical use under the following conditions:
i.
Whenever quality assurance check measurements indicate
that the radiation output differs by more than 5 percent from
the value obtained at the last full calibration and the
difference cannot be easily discerned; and
ii.
Following any component replacement, major repair, or
modification of components that could significantly affect
the characteristics of the radiation beam.
iv.
Notwithstanding the requirements of 1.15.7(20)(a)(iii):
i.
Full calibration of therapeutic radiation machines with
multi-energy and/or multi-mode capabilities is required only
for those modes and/or energies that are not within their
acceptable range; and
ii.
If the repair, replacement or modification does not affect all
modes and/or energies, full calibration shall be performed on
the affected mode/energy that is in most frequent clinical use
at the facility. The remaining energies/modes may be
validated with quality assurance check procedures against
the criteria in 1.15.7(20)(a)(iii)(i).
b.
To satisfy the requirement of 1.15.7(1)(a), full calibration shall include all
measurements required for annual calibration by American Association of
Physicists in Medicine (AAPM) Report 13, "Physical Aspects of Quality
Assurance in Radiation Therapy";
c.
The registrant shall use the dosimetry system described in 1.15.4(3) to
measure the radiation output for one set of exposure conditions. The
remaining radiation measurements required in 1.15.7(20)(b) may then be
made using a dosimetry system that indicates relative dose rates; and
d.
The registrant shall maintain a record of each calibration for the duration of
the registration. The record shall include the date of the calibration, the
manufacturer's name, model number, and serial number for the therapeutic
radiation machine, the model numbers and serial numbers of the
instruments used to calibrate the therapeutic radiation machine, and the
signature of the Radiation Therapy Physicist responsible for performing the
calibration.
21.
Periodic Quality Assurance Checks.
a.
Periodic quality assurance checks shall be performed on all therapeutic
radiation machines subject to 1.15.7 at intervals not to exceed 1 week;
b.
To satisfy the requirement of 1.15.7(21)(a), quality assurance checks shall
include determination of all parameters for periodic quality assurance
checks contained in American Association of Physicists in Medicine
(AAPM) Report 13, "Physical Aspects of Quality Assurance in Radiation
Therapy";
c.
The registrant shall use a dosimetry system which has been intercompared
within the previous 6 months with the dosimetry system described in
1.15.4(3) to make the periodic quality assurance checks required in
1.15.7(21)(b);
d.
The registrant shall perform periodic quality assurance checks required by
1.15.7(21)(a) in accordance with procedures established by the Radiation
Therapy Physicist;
e.
The registrant shall review the results of each periodic radiation output
check according to the following procedures:
i.
The authorized user and Radiation Therapy Physicist shall be
immediately notified if any parameter is not within its acceptable
range. The therapeutic radiation machine shall not be made
available for subsequent medical use until the Radiation Therapy
Physicist has determined that all parameters are within their
acceptable range;
ii.
If all quality assurance check parameters appear to be within their
acceptable range, the quality assurance check shall be reviewed and
signed by either the authorized user or Radiation Therapy Physicist
within 3 treatment days; and
iii.
The Radiation Therapy Physicist shall review and sign the results of
each radiation output quality assurance check at intervals not to
exceed 1 month;
f.
Therapeutic radiation machines subject to 1.15.7 shall have safety quality
assurance checks of each radiation therapy facility performed at intervals
not to exceed 1 week;
g.
To satisfy the requirement of 1.15.7(21)(f), safety quality assurance checks
shall ensure proper operation of:
i.
Electrical interlocks at each external beam radiation therapy room
entrance;
ii.
Proper operation of the "BEAM-ON", interrupt and termination
switches;
iii.
Beam condition indicator lights on the access doors, control console,
and in the radiation therapy room;
iv.
Viewing systems;
v.
Electrically operated treatment room door(s) from inside and outside
the treatment room; and
vi.
At least one emergency power cutoff switch. If more than one
emergency power cutoff switch is installed and not all switches are
tested at once, each switch shall be tested on a rotating basis. Safety
quality assurance checks of the emergency power cutoff switches
may be conducted at the end of the treatment day in order to
minimize possible stability problems with the therapeutic radiation
machine.
h.
The registrant shall promptly repair any system identified in 1.15.7(21)(g)
that is not operating properly; and
i.
A registrant shall maintain a record of each quality assurance check required
by 1.15.7(21)(a) and 1.15.7(21)(g) for 3 years. The record shall include the
date of the quality assurance check, the manufacturer's name, model
number, and serial number for the therapeutic radiation machine, the
manufacturer's name, model number and serial number of the instrument(s)
used to measure the radiation output of the therapeutic radiation machine,
and the signature of the individual who performed the periodic quality
assurance check.