15 MAC Pt. 21, R. 1.6.3
General and Administrative Requirements
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.6.3
General and Administrative Requirements.
1. Radiation Safety Requirements. The registrant shall be responsible for directing the
operation of the x-ray system(s) under his administrative control. The registrant or
the registrant's agent shall assure that the requirements of these regulations are met
in the operation of the x-ray system(s).
a.
An x-ray system which does not meet the provisions of these regulations
shall not be operated for diagnostic purposes (if so directed by the Agency).
b.
Individuals who will be operating the x-ray systems shall be adequately
instructed in the safe operating procedures and be competent in the safe use
of the equipment. See Appendix A for a list of subject matters pertinent to
this requirement. The Agency may use interview, observation and/or
testing to determine compliance.
c.
A chart shall be provided in the vicinity of the diagnostic x-ray system's
control panel which specifies, for all examinations performed with that
system, the following information:
i.
patient's body part and anatomical size, or body part thickness, or
age (for pediatrics), versus technique factors to be utilized;
ii.
type and size of the film or screen-film combination to be used;
iii.
type and focal distance of the grid to be used, if any;
iv.
source to image receptor distance to be used (except for dental
intraoral radiography); and
v.
type and location of placement of patient shielding (i.e., gonad, etc.)
to be used.
d.
The registrant of a facility shall establish and make available to x-ray
operators written safety procedures, including patient holding and any
restrictions of the operating technique required for the safe operation of the
particular x-ray system. The operator shall be able to demonstrate
familiarity with these procedures.
e.
Except for human patients who cannot be moved out of the room, only the
staff and ancillary personnel, or other persons, required for the medical
procedure or training shall be in the room during the radiographic exposure.
Other than the patient being examined:
i.
All individuals shall be positioned such that no part of the body will
be struck by the useful beam unless protected by not less than 0.5
millimeter lead equivalent material.
ii.
The x-ray operator, other staff, ancillary personnel, and other
persons required for the medical procedure shall be protected from
the direct scatter radiation by protective aprons or whole body
protective barriers of not less than 0.25 millimeter lead equivalent
material.
iii.
Human patients who cannot be removed from the room shall be
protected from the direct scatter radiation by whole body protective
barriers of not less than 0.25 millimeter lead equivalent material, or
shall be so positioned that the nearest portion of the body is at least
2 meters from both the tube head and the nearest edge of the image
receptor.
f. Gonad shielding of not less than 0.5 millimeter lead equivalent material shall be
used for human patients, who have not passed the reproductive age, during
radiographic procedures in which the gonads are in the useful beam, except
for cases in which this would interfere with the diagnostic procedure.
g.
Individuals shall not be exposed to the useful beam except for healing arts
purposes and unless such exposure has been authorized by a licensed
practitioner of the healing arts. This provision specifically prohibits
deliberate exposure for the following purposes:
i.
exposure of an individual for training, demonstration, or other
non-healing-arts purposes; and
ii.
exposure of an individual for the purpose of healing arts screening
except as authorized by 1.6.3(1)(k).
h.
When a patient or film must be provided with auxiliary support during a
radiation exposure:
i.
mechanical holding devices shall be used when the technique
permits. The written safety procedures, required by 1.6.3(1)(d),
shall list individual projections where holding devices cannot be
utilized;
ii.
written safety procedures, as required by 1.6.3(1)(d), shall indicate
the requirements for selecting a holder and the procedure the holder
shall follow;
iii.
the human holder shall be instructed in personal radiation safety and
protected as required by 1.6.3(1)(e);
iv.
no individual shall be used routinely to hold film or patients;
v.
in those cases where the patient must hold the film, except during
intraoral examinations, any portion of the body other than the area
of clinical interest struck by the useful beam shall be protected by
not less than 0.5 millimeter lead equivalent material;
vi.
when an animal must be held by an individual during an exposure,
that individual shall be protected with appropriate shielding devices,
such as leaded aprons and gloves, and shall be positioned such that
no part of his or her body shall be struck by the useful beam; and
vii.
each facility must have leaded aprons and gloves available in
sufficient numbers to provide protection to all personnel who are
involved with x-ray operations and who are otherwise not shielded.
i. Procedures and auxiliary equipment designed to minimize patient and personnel
exposure commensurate with the needed diagnostic information shall be
utilized.
i.
The speed of the screen and film combinations used shall be the
fastest speed consistent with the diagnostic objective of the
examinations. Film cassettes without intensifying screens shall not
be used for any routine diagnostic radiological imaging, with the
exception of standard film packets for intraoral use in dental
radiography.
ii.
The radiation exposure to the patient shall be the minimum exposure
required to produce images of good diagnostic quality.
iii.
Portable or mobile x-ray equipment shall be used only for
examinations where it is impractical to transfer the patient(s) to a
stationary x-ray installation.
iv.
X-ray systems subject to 1.6.6 shall not be utilized in procedures
where the source to patient distance is less than 30 centimeters
except for veterinary systems.
v.
If grids are used between the patient and the image receptor to
decrease scatter to the film and improve contrast the grid shall:
vi.
Be positioned properly, i.e., tube side facing the right direction, and
grid centered to the central ray.
vii.
If of the focused type, be of the proper focal distance for the SIDs
being used.
j. All individuals who are associated with the operation of an x-ray system are
subject to the requirements of 1.4.5 and 1.4.7 of these regulations. In
addition:
i.
When protective clothing or devices are worn on portions of the
body and a personnel monitoring device(s) is required, at least one
such monitoring device shall be utilized as follows:
i.
When an apron is worn, the monitoring device shall be worn
at the collar outside of the apron.
ii.
The dose to the whole body based on the maximum dose
attributed to the most critical organ shall be recorded in the
reports required by 1.4.6 of these regulations. If more than
one device is used and a record is made of the data, each dose
shall be identified with the area where the device was worn
on the body.
ii.
Exposure of a personnel monitoring device to deceptively indicate a
dose delivered to an individual is prohibited.
k.
Healing Arts Screening. Any person proposing to conduct a healing arts
screening program shall not initiate such a program without prior approval
of the Agency. When requesting such approval, that person shall submit
the information outlined in Appendix B of this Subchapter. If any
information submitted to the Agency becomes invalid or outdated, the
Agency shall be immediately notified.
l. Information and Maintenance Record and Associated Information. The
registrant shall maintain the following information for each x-ray system
for inspection by the Agency:
i.
model and serial numbers of all major components, and user's
manuals for those components;
ii.
tube rating charts and cooling curves;
iii.
records of surveys, calibrations, maintenance, and modifications
performed on the x-ray system(s); and
iv.
a copy of all correspondence with this Agency regarding that x-ray
system.
m.
X-Ray Log. Each facility shall maintain an x-ray log containing the
patient's name, the type of examinations, and the dates the examinations
were performed. When the patient or film must be provided with human
auxiliary support, the name of the human holder shall be recorded.
n.
A sign shall be posted in a conspicuous area so as to be easily seen by the
patient to the effect that if there is a pregnancy or the possibility of a
pregnancy, the physician should be notified. Prescription of x-ray
examinations of pregnant or possibly pregnant patients shall assure that
medical consideration has been given to possible fetal exposure and
appropriate protective measures are taken.
2. X-ray Film Processing Facilities and Practices. Each installation using a
radiographic x-ray system and using analog image receptors (radiographic film)
shall have available suitable equipment for handling and processing radiographic
film in accordance with the following provisions:
a.
Manually developed film:
i.
Processing tanks shall be constructed of mechanically rigid,
corrosion resistant material; and
ii.
The temperature of solutions in the tanks shall be maintained within
the range of 60ºF to 80ºF (16ºC to 27ºC). Film shall be developed
in
accordance
with
the
time-temperature
relationships
recommended by the film manufacturer, or, in the absence of such
recommendations, with the time-temperature chart below:
Time-Temperature Chart
Thermometer
Reading
(Degrees)
Minimum
Developing
Time
(Minutes)
oC
oF
26.7
26.1
25.6
2 ½
25.0
2 ½
24.4
23.9
23.3
3 ½
22.8
3 ½
22.2
21.7
21.1
4 ½
20.6
4 ½
20.0
19.4
5 ½
18.9
5 ½
18.3
17.8
6 ½
17.2
16.7
16.1
8 ½
15.6
9 ½
iii.
Devices shall be utilized which will:
i.
Indicate the actual temperature of the developer; and
ii.
Signal the passage of a preset time appropriate to the
developing time required.
b.
Automatic Processors and Other Closed Processing Systems:
i.
Films shall be developed in accordance with the time-temperature
relationships recommended by the film manufacturer; in the absence
of such recommendations, the film shall be developed using the
chart below:
Developer
Temperature
(Degrees)
Minimum
Immersion
Time:*
(Seconds)
oC
oF
35.5
34.5
33.5
31.5
30.5
29.5
*Immersion time only, no crossover time included.
ii.
The specified developer temperature and immersion time shall be
posted in the darkroom or on the automatic processor.
c.
Other Requirements
i.
Pass boxes, if provided, shall be so constructed as to exclude light
from the darkroom when cassettes are placed in or removed from
the boxes, and shall incorporate adequate shielding from stray
radiation to prevent exposure of undeveloped film.
ii.
The darkroom shall be light tight and use proper safelighting such
that any film type in use exposed in a cassette to x-radiation
sufficient to produce an optical density from 1 to 2 when processed
shall not suffer an increase in density greater than 0.1 (0.02 for
mammography) when exposed in the darkroom for 2 minutes with
all safelights on. If used, daylight film handling boxes shall
preclude fogging of the film.
iii.
Darkrooms typically used by more than one individual shall be
provided a positive method to prevent accidental entry while
undeveloped films are being handled or processed.
iv.
Film shall be stored in a cool, dry place and shall be protected from
exposure to stray radiation. Film in open packages shall be stored
in a light tight container.
i.
Film cassettes and intensifying screens shall be inspected
periodically and shall be cleaned and replaced as necessary
to best assure radiographs of good diagnostic quality.
ii.
Outdated x-ray film shall not be used for human diagnostic
radiographs, unless the film has been stored in accordance
with the manufacturer's recommendations and a sample of
the film passes a sensitometric test for normal ranges of base
plus fog and speed.
iii.
Film developing solutions shall be prepared in accordance
with the directions given by the manufacturer, and shall be
maintained in strength by replenishment or renewal so that
full development is accomplished within the time specified
by the manufacturer.