15 MAC Pt. 21, R. 1.6.4
General Requirements for all Diagnostic X-Ray Systems
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.6.4
General Requirements for all Diagnostic X-Ray Systems. In addition to other
requirements of this section, all diagnostic x-ray systems shall meet the following
requirements:
1. Warning Label. The control panel containing the main power switch shall bear the
warning statement, legible and accessible to view: "WARNING: This x-ray unit
may be dangerous to patient and operator unless safe exposure factors and operating
instructions are observed."
2. Battery Charge Indicator. On battery-powered x-ray generators, visual means shall
be provided on the control panel to indicate whether the battery is in a state of
charge adequate for proper operation.
3. Leakage Radiation from the Diagnostic Source Assembly. The leakage radiation
from the diagnostic source assembly measured at a distance of 1 meter in any
direction from the source shall not exceed 100 milliroentgens (25.8 µC/kg) in 1
hour when the x-ray tube is operated at its leakage technique factors. Compliance
shall be determined by measurements averaged over an area of 100 square
centimeters with no linear dimension greater than 20 centimeters. If leakage
technique factors cannot be set on the control panel, then compliance shall be
determined by measuring leakage at maximum kVp and an appropriate mAs.
4. Radiation from Components Other Than the Diagnostic Source Assembly. The
radiation emitted by a component other than the diagnostic source assembly shall
not exceed 2 milliroentgens (0.516 µC/kg) in 1 hour at 5 centimeters from any
accessible surface of the component when it is operated in an assembled x-ray
system under any conditions for which it was designed. Compliance shall be
determined by measurements averaged over an area of 100 square centimeters with
no linear dimension greater than 20 centimeters.
5. Beam Quality.
a.
Half-value Layer.
i.
The half-value layer of the useful beam for a given x-ray tube
potential shall not be less than the values shown in Table I. If it is
necessary to determine such half-value layer at an x-ray tube
potential which is not listed in Table I, linear interpolation or
extrapolation may be made.
TABLE I
X-ray tube voltage
(kilovolt peak)
Minimum HVL (mm of Al)
Designed Operating
Range
Measured
Operating
Potential
Dental
Systems
Medical
X-ray Systems
Manufactured
before
06/10/2006
Medical
X-ray Systems
Manufactured
after
06/10/2006
Below 50
1.5
0.3
0.3
1.5
0.4
0.4
1.5
0.5
0.5
50 to 70
1.5
1.3
1.3
1.5
1.5
1.5
1.5
2.1
1.8
Above 70
2.1
2.1
2.5
2.3
2.3
2.9
2.5
2.5
3.2
2.7
2.7
3.6
3.0
3.0
3.9
3.2
3.2
4.3
3.5
3.5
4.7
3.8
3.8
5.0
4.1
4.1
5.4
ii.
For capacitor energy storage equipment, compliance with the
requirements of 1.6.4(5) shall be determined with the maximum
quantity of charge per exposure. This will be deemed to have been
met if a mAs of 5-10 has been used.
iii.
The required minimal half-value-layer of the useful beam shall
include the filtration contributed by all materials which are
permanently between the source and the patient.
b. Filtration Controls. For x-ray systems which have variable kVp and variable
filtration for the useful beam, a device shall link the kVp selector with the
filter(s) and shall prevent an exposure unless the minimum amount of
filtration necessary to produce the HVL required by 1.6.4(5)(a) is in the
useful beam for the given kVp which has been selected.
6. Multiple Tubes. Where two or more radiographic tubes are controlled by one
exposure switch, the tube or tubes which have been selected shall be clearly
indicated prior to initiation of the exposure. This indication shall be both on the
x-ray control panel and at or near the tube housing assembly which has been
selected.
7. Mechanical Support of Tube Head. The tube housing assembly supports shall be
adjusted such that the tube housing assembly will remain stable during an exposure
unless tube housing movement is a designed function of the x-ray system.
8. Technique Indicators.
a. The technique factors to be used during an exposure shall be indicated before
the exposure begins. If automatic exposure controls are used, the technique
factors which are set prior to the exposure shall be indicated.
b. The requirements of 1.6.4(8)(a) may be met by permanent markings on
equipment having fixed technique factors. Indication of technique factors shall
be visible from the operator's position except in the case of spot films made by
the fluoroscopist.
9. Locks. All position locking, holding, and centering devices on x-ray system
components shall function as intended.