15 MAC Pt. 21, R. 1.6.8
Veterinary Medicine Radiographic Installations
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.6.8
Veterinary Medicine Radiographic Installations.
1. Equipment.
a.
The protective tube housing shall be equivalent to the requirements of
1.6.4(3).
b.
Diaphragms or cones shall be provided for collimating the useful beam to
the area of clinical interest and shall provide the same degree of protection
as is required of the housing.
c.
The total filtration permanently in the useful beam shall not be less than 0.5
millimeters aluminum equivalent for machines operating up to 50 kVp, 1.5
millimeters aluminum equivalent for machines operating between 50 and
70 kVp, and 2.5 millimeters aluminum equivalent for machines operating
above 70 kVp.
d.
A device shall be provided to terminate the exposure after a preset time or
exposure.
e.
A dead-man type of exposure switch shall be provided, together with an
electrical cord of sufficient length, so that the operator can stand out of the
useful beam and at least 6 feet (1.83 m) from the animal during all x-ray
exposures.
2. Structural Shielding. All wall, ceiling, and floor areas shall be equivalent to or
provided with applicable protective barriers to assure compliance with 1.4.5 of
these regulations.
3. Operating Procedures.
a.
The operator shall stand well away from the useful beam and the animal
during radiographic exposures.
b.
No individual other than the operator shall be in the x-ray room while
exposures are being made unless such individual's assistance is required.
c.
When an animal must be held in position during radiography, mechanical
supporting or restraining devices should be used. If the animal must be held
by an individual, that individual shall be protected with appropriate
shielding devices, such as protective gloves and apron, and he shall be so
positioned that no part of his body will be struck by the useful beam. The
exposure of any individual used for this purpose shall be monitored.