15 MAC Pt. 21, R. 1.6.9
Computed Tomography X-Ray Systems
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.6.9
Computed Tomography X-Ray Systems.
1. Definitions. In addition to the definitions provided in 1.1.2 and 1.6.2 of these
regulations, the following definitions shall be applicable to 1.6.9:
a.
"Computed tomography dose index" means the integral from -7T to +7T of
the dose profile along a line perpendicular to the tomographic plane divided
by the product of the nominal tomographic section thickness and the number
of tomograms produced in a single scan, that is:
where:
z
= Position along a line perpendicular to the tomographic plane.
D(z) = Dose at position z.
T
= Nominal tomographic section thickness.
n
= Number of tomograms produced in a single scan.
This definition assumes that the dose profile is centered around z = 0 and
that, for a multiple tomogram system, the scan increment between adjacent
scans is nT.
b.
"Contrast scale" means the change in the linear attenuation coefficient per
CTN relative to water, that is:
where:
μx
=
Linear attenuation coefficient of the material of interest.
μw
=
Linear attenuation coefficient of water.
(CTN)x =
CTN of the material of interest.
(CTN)w =
CTN of water.
c.
"CS" (See "Contrast scale").
d.
"CT conditions of operation" means all selectable parameters governing the
operation of a CT x-ray system including, but not limited to, nominal
tomographic section thickness, filtration, and the technique factors as
defined in 1.6.2.
e.
"CTDI" (See "Computed tomography dose index").
f. "CT gantry" means the tube housing assemblies, beam-limiting devices, detectors,
and the supporting structures and frames which hold these components.
g.
"CTN" (See "CT number").
h.
"CT number" means the number used to represent the x-ray attenuation
associated with each elemental area of the CT image.
where:
k
=
A constant5
μx =
Linear attenuation coefficient of the material of interest.
μw =
Linear attenuation coefficient of water.
i. "Dose profile" means the dose as a function of position along a line.
j. "Elemental area" means the smallest area within a tomogram for which the x-ray
attenuation properties of a body are depicted. (See also "Picture element").
5 The constant has a normal value of 1,000 when the Houndsfield scale of CTN is used.
k.
"Multiple tomogram system" means a computed tomography x-ray system
which obtains x-ray transmission data simultaneously during a single scan
to produce more than one tomogram.
l. "Noise" means the standard deviation of the fluctuations in CTN expressed as a
percentage of the attenuation coefficient of water. Its estimate (Sn) is
calculated using the following expression:
where:
CS =
Contract scale.
μw =
Linear attenuation coefficient of water.
s
=
Estimated standard deviation of the CTN of picture elements in a
specified area of the CT image.
m.
"Nominal tomographic section thickness" means the full width at
half-maximum of the sensitivity profile taken at the center of the
cross-sectional volume over which x-ray transmission data are collected.
n.
"Picture element" means an elemental area of a tomogram.
o.
"Reference plane" means a plane which is displaced from and parallel to the
tomographic plane.
p.
"Scan" means the complete process of collecting x-ray transmission data for
the production of a tomogram. Data can be collected simultaneously during
a single scan for the production of one or more tomograms.
q.
"Scan increment" means the amount of relative displacement of the patient
with respect to the CT x-ray system between successive scans measured
along the direction of such displacement.
r. "Scan sequence" means a preselected set of two or more scans performed
consecutively under preselected CT conditions of operation.
s.
"Scan time" means the period of time between the beginning and end of
x-ray transmission data accumulation for a single scan.
t. "Single tomogram system" means a CT x-ray system which obtains x-ray
transmission data during a scan to produce a single tomogram.
u.
"Tomographic plane" means that geometric plane which is identified as
corresponding to the output tomogram.
v.
"Tomographic section" means the volume of an object whose x-ray
attenuation properties are imaged in a tomogram.
2. Requirement for Equipment.
a.
Termination of Exposure.
i.
Means shall be provided to terminate the x-ray exposure
automatically by either de-energizing the x-ray source or
shuttering the x-ray beam in the event of equipment failure
affecting data collection. Such termination shall occur within an
interval that limits the total scan time to no more than 110 percent
of its preset value through the use of either a backup timer or devices
which monitor equipment function.
ii.
A visible signal shall indicate when the x-ray exposure has been
terminated through the means required by 1.6.9(2)(a)(i).
iii.
The operator shall be able to terminate the x-ray exposure at any
time during a scan, or series of scans under CT x-ray system control,
of greater than one-half second duration.
b.
Tomographic Plane Indication and Alignment.
i.
For any single tomogram system, means shall be provided to permit
visual determination of the tomographic plane or a reference plan
offset from the tomographic plane.
ii.
For any multiple tomogram system, means shall be provided to
permit visual determination of the location of a reference plane.
This reference plane can be offset from the location of the
tomographic planes.
iii.
If a device using a light source is used to satisfy 1.6.9(2)(b)(i) or (ii),
the light source shall provide illumination levels sufficient to permit
visual determination of the location of the tomographic plane or
reference plane under ambient light conditions of up to 500 lux.
c.
Beam-On and Shutter Status Indicators and Control Switches.
i.
The CT x-ray control and gantry shall provide visual indication
whenever x-rays are produced and, if applicable, whether the shutter
is open or closed.
ii.
Each emergency button or switch shall be clearly labeled as to its
function.
d.
Indication of CT Conditions of Operation. The CT x-ray system shall be
designed such that the CT conditions of operation to be used during a scan
or a scan sequence shall be indicated prior to the initiation of a scan or a
scan sequence. On equipment having all or some of these conditions of
operation at fixed values, this requirement may be met by permanent
markings. Indication of CT conditions of operation shall be visible from
any position from which scan initiation is possible.
e.
Extraneous Radiation. When data are not being collected for image
production, the radiation adjacent to the tube port shall not exceed that
permitted by 1.6.4(3).
f. Maximum Surface CTDI Identification. The angular position where the
maximum surface CTDI occurs shall be identified to allow for reproducible
positioning of a CT dosimetry phantom.
g.
Additional Requirements Applicable to CT X-Ray Systems Containing
a Gantry Manufactured After September 3, 1985.
i.
The total error in the indicated location of the tomographic plane or
reference plane shall not exceed 5 millimeters.
ii.
If the x-ray production period is less than one-half second, the
indication of x-ray production shall be actuated for at least one-half
second. Indicators at or near the gantry shall be discernible from
any point external to the patient opening where insertion of any part
of the human body into the primary beam is possible.
iii.
The deviation of indicated scan increment versus actual increment
shall not exceed plus or minus 1 millimeter with any mass from 0 to
100 kilograms resting on the support device. The patient support
device shall be incremented from a typical starting position to the
maximum incremented distance or 30 centimeters, whichever is
less, and then returned to the starting position. Measurement of
actual versus indicated scan increment may be taken anywhere
along this travel.
iv.
Premature termination of the x-ray exposure by the operator shall
necessitate resetting of the CT conditions of operation prior to the
initiation of another scan.
3. Facility Design Requirements.
a.
Aural Communication. Provision shall be made for two-way aural
communication between the patient and the operator at the control panel.
b.
Viewing Systems.
i.
Windows, mirrors, closed-circuit television, or an equivalent shall
be provided to permit continuous observation of the patient during
irradiation and shall be so located that the operator can observe the
patient from the control panel.
ii.
When the primary viewing system is by electronic means, an
alternate viewing system (which may be electronic) shall be
available for use in the event of failure of the primary viewing
system.
4. Surveys, Measurements, Spot Checks and Operating Procedures.
a.
Surveys.
i.
All CT x-ray systems installed after the effective date of these
regulations and those systems not previously surveyed shall have a
survey made by, or under the direction of, a qualified expert. In
addition, such surveys shall be done after any change in the facility
or equipment which might cause a significant increase in radiation
hazard.
ii.
The registrant shall obtain a written report of the survey from the
qualified expert, and a copy of the report shall be made available to
the Agency upon request.
b.
Radiation Measurements.
i.
The measurements of the radiation output of the CT x-ray system
shall be performed by, or under the direction of, a qualified expert.
ii.
The measurement of the radiation output of the CT x-ray system
shall be performed annually and after any change or replacement of
components which could cause a change in the radiation output.
iii.
The measurement of the radiation output of the CT x-ray system
shall be performed with a calibrated dosimetry system. The
calibration of such system shall be traceable to a national standard.
The dosimetry system shall have been calibrated within the
preceding 2 years.
iv.
CT dosimetry phantom(s) shall be used in determining the radiation
output of a CT x-ray system. Such phantom(s) shall meet the
following specifications and conditions of use:
i.
CT dosimetry phantom(s) shall be right circular cylinders of
polymethyl-methacrylate of density 1.19 plus or minus 0.01
grams per cubic centimeter. The phantom(s) shall be at least
14 centimeters in length and shall have diameters of 32.0
centimeters for testing CT x-ray systems designed to image
any section of the body and 16.0 centimeters for systems
designed to image the head or for whole body scanners
operated in the head scanning mode.
ii.
CT dosimetry phantom(s) shall provide means for the
placement of a dosimeter(s) along the axis of rotation and
along a line parallel to the axis of rotation 1.0 centimeter
from the outer surface and within the phantom. Means for
the placement of dosimeters or alignment devices at other
locations may be provided.
iii.
Any effects on the doses measured to the removal of
phantom material to accommodate dosimeters shall be
accounted for through appropriate corrections to the reported
data or included in the statement of maximum deviation for
the values obtained using the phantom.
iv.
All dose measurements shall be performed with the CT
dosimetry phantom placed on the patient couch or support
device without additional attenuation materials present.
v.
The measurement of the radiation output shall be required
for each type of head, body, or whole-body scan performed
at the facility.
vi.
Measurement of the radiation output shall meet the
following requirements:
i.
The dose profile along the center axis of the CT
dosimetry phantom for the minimum, maximum, and
midrange values of the nominal tomographic section
thickness used by the registrant shall be measurable.
Where less than three nominal tomographic
thicknesses can be selected, the dose profile
determination shall be performed for each available
nominal tomographic section thickness.
ii.
The CTDI6 along the two axes specified in
1.6.9(4)(b)(iv)(ii) shall be measured. The CT
dosimetry phantom shall be oriented so that the
measurement point 1.0 centimeter from the outer
surface and within the phantom is in the same
angular position within the gantry as the point of
6For the purpose of determining the CTDI, the manufacturer's statements as to the nominal
tomographic section thickness for that particular system may be utilized.
maximum surface CTDI identified. The CT
conditions of operation shall correspond to typical
values used by the registrant.
iii.
The spot checks specified in 1.6.9(4)(c) shall be
made.
vii.
Procedures for the measurement of radiation output shall be
in writing. Records of measurements performed shall be
maintained for inspection by the Agency.
c.
Spot Checks.
i.
The spot-check procedures shall be in writing and shall have been
developed by a qualified expert.
ii.
The spot-check procedures shall incorporate the use of a CT
dosimetry phantom which has a capability of providing an
indication of contrast scale, noise, nominal tomographic section
thickness, the resolution capability of the system for low and high
contrast objects, and measuring the mean CTN for water or other
reference material.
iii.
All spot checks shall be included in the measurement required by
1.6.9(4)(b) and at time intervals and under system conditions
specified by a qualified expert.
iv.
Spot checks shall include acquisition of images obtained with the
CT dosimetry phantom(s) using the same processing mode and CT
conditions of operation as are used to perform measurements
required by 1.6.9(4)(b). The images shall be retained, until a new
measurement is performed, in two forms as follows:
i.
photographic copies of the images obtained from the image
display device; and
ii.
images stored in digital form on a storage medium
compatible with the CT x-ray system.
v.
Written records of the spot checks performed shall be maintained
for inspection by the Agency.
d.
Operating Procedures.
i.
The CT x-ray system shall not be operated except by an individual
who has been specifically trained in its operation.
ii.
Information shall be available at the control panel regarding the
operation of the system and measurements of radiation output. Such
information shall include the following:
i.
dates of the latest measurements and spot checks and the
location within the facility where the results of those tests
may be obtained;
ii.
instructions on the use of the CT dosimetry phantom(s)
including a schedule of spot checks appropriate for the
system, allowable variations for the indicated parameters,
and the results of at least the most recent spot checks
conducted on the system;
iii.
the distance in millimeters between the tomographic plane
and the reference plane if a reference plane is utilized; and
iv.
a current technique chart available at the control panel which
specifies for each routine examination the CT conditions of
operation and the number of scans per examination.
iii.
If the measurement or spot check of the CT x-ray system identifies
that a system operating parameter has exceeded a tolerance
established by the qualified expert, use of the CT x-ray system on
patients shall be limited to those permitted by established written
instructions of the qualified expert.