15 MAC Pt. 21, R. 1.8.3
Equipment Requirements
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.8.3
Equipment Requirements.
1. Safety Device. A device which prevents the entry of any portion of an individual's
body into the primary x-ray beam path or which causes the beam to be shut off
upon entry into its path shall be provided on all open- beam configurations. A
registrant may apply to the Agency for an exemption from the requirements of a
safety device. Such application shall include:
a.
a description of the various safety devices that have been evaluated;
b.
the reason each of these devices cannot be used; and
c.
a description of the alternative methods that will be employed to minimize
the possibility of an accidental exposure, including procedures to assure
that operators and others in the area will be informed of the absence of
safety devices.
2. Warning Devices
a.
Open-beam configurations shall be provided with a readily discernible
indication of:
i.
x-ray tube "on-off" status located near the radiation source
housing, if the primary beam is controlled in this manner; and/or
ii.
shutter "open-closed" status located near each port on the radiation
source housing, if the primary beam is controlled in this manner.
b.
An easily visible warning light labeled with the words "X-RAY ON", or
words having a similar intent, shall be located:
i.
near any switch that energizes an x-ray tube and shall be
illuminated only when the tube is energized; or
ii.
in the case of a radioactive source, near any switch that opens a
housing shutter and shall be illuminated only when the shutter is
open.
c.
Warning devices shall be labeled so that their purpose is easily
identified. On equipment i n st a l l e d aft er June 30, 1978,
warning devices shall have fail-safe characteristics.
3. Ports. Unused ports on radiation source housings shall be secured in the closed
position in a manner which will prevent casual opening.
4. Labeling. All analytical x-ray equipment shall be labeled with a readily discernible
sign or signs bearing the radiation symbol and the words:
a. "CAUTION - HIGH INTENSITY X-RAY BEAM", or words having
a similar intent, on the x-ray source housing; and
b. "CAUTION RADIATION - THIS EQUIPMENT PRODUCES
RADIATION WHEN ENERGIZED", or words having a similar
intent, near any switch that energizes an x-ray tube if the radiation source
is an x-ray tube; or
c. "CAUTION - RADIOACTIVE MATERIAL", or words having a
similar intent, on the source housing in accordance with 1 . 4.29 of
these regulations if the radiation source is a radionuclide.
5.
Shutters. On open-beam configurations installed after January 12, 1980, each
port on the radiation source housing shall be equipped with a shutter that cannot
be opened unless a collimator or a coupling has been connected to the
port.
6.
Radiation Source Housing. Each radiation source housing shall be subject to
the following requirements:
a.
Each x-ray tube housing shall be equipped with an interlock that shuts off
the tube if it is removed from the radiation source housing or if the housing
is disassembled.
b.
Each radioactive source housing or port cover or each x-ray tube
housing shall be so constructed that, with all shutters closed, the radiation
measured at a distance of 5 centimeters from its surface is not capable
of producing a dose in excess of 2.5 millirems (0.025 mSv) in one
hour. For systems utilizing x-ray tubes, this limit shall be met at any
specified tube rating.
7.
Generator C a b i n e t . Each x -ray generat or s h a l l b e supplied w i t h
a protective cabinet which limits leakage radiation measured at a distance of 5
centimeters from its surface such that it is not capable of producing a dose in
excess of 0.25 millirem (2.5 ยตSv) in one hour.