15 MAC Pt. 21, R. 1.9.10
Operating Procedures
Cite as 15 Miss. Admin. Code Pt. 21, R. 1.9.10
Operating Procedures.
1.
Particle accelerators, when not in operation, shall be secured to prevent
unauthorized use.
2.
The safety interlock system shall not be used to turn off the accelerator beam
except in an emergency.
3.
All safety and warning devices, including interlocks, shall be checked for proper
operation at intervals not to exceed three months. Results of such tests shall be
maintained at the accelerator facility for inspection by the Agency.
4.
Electrical circuit diagrams of the accelerator and the associated safety interlock
systems shall be kept current and maintained for inspection by the Agency and
shall be available to the operator at each accelerator facility.
5.
If, for any reason, it is necessary to intentionally bypass a safety interlock or
interlocks, such action shall be:
a.
authorized by the radiation safety committee and/or radiation safety
officer;
b.
recorded in a permanent log and a notice posted at the accelerator
control console; and
c.
terminated as soon as possible.
6.
A copy of the current operating and the emergency procedures shall be maintained
at the accelerator control panel.