NDAC 33.1-10-06-04
General requirements for all diagnostic x-ray systems
Cite as N.D. Admin. Code ยง 33.1-10-06-04
In addition to other requirements of this chapter, all diagnostic x-ray systems shall meet the
following requirements:
1.
Warning label. The control panel containing the main power switch shall bear the warning
statement, legible and accessible to view: "WARNING: This x-ray unit may be dangerous to
patient and operator unless safe exposure factors and operating instructions are observed."
2.
Battery charge indicator. On battery-powered x-ray generators, visual means shall be
provided on the control panel to indicate whether the battery is in a state of charge adequate
for proper operation.
3.
Leakage radiation from the diagnostic source assembly. The leakage radiation from the
diagnostic source assembly measured at a distance of one meter in any direction from the
source shall not exceed one hundred milliroentgens in one hour when the x-ray tube is
operated at its leakage technique factors. Compliance shall be determined by measurements
averaged over an area of one hundred square centimeters with no linear dimension greater
than twenty centimeters.
4.
Radiation from components other than the diagnostic source assembly. The radiation
emitted by a component other than the diagnostic source assembly shall not exceed
two milliroentgens in one hour at five centimeters from any accessible surface of the
component when it is operated in an assembled x-ray system under any conditions for which it
was designed. Compliance shall be determined by measurements averaged over an area of
one hundred square centimeters with no linear dimension greater than twenty centimeters.
5.
Beam quality.
a.
Half-value layer.
(1)
The half-value layer (HVL) of the useful beam for a given x-ray tube potential shall
not be less than the values shown in table I. If it is necessary to determine such half
value layer at an x-ray tube potential which is not listed in table I, linear interpolation
or extrapolation may be made.
TABLE I
Half-Value Layer in Millimeters Aluminum
Design
Operating
Range
(Kilovolts Peak)
Measured
Potential
(Kilovolts Peak)
Dental Intraoral
Manufactured
Before Aug. 1,
1974, and on
or After Dec. 1,
1980
Diagnostic
X-Ray Systems
Manufactured
Prior to
June 10, 2006
Diagnostic
X-Ray Systems
Manufactured
on or After
June 10, 2006
Below 51
30
N/A
0.3
0.3
40
N/A
0.4
0.4
50
1.5
0.5
0.5
51 to 70
51
1.5
1.2
1.3
60
1.5
1.3
1.5
70
1.5
1.5
1.8
Above 70
71
2.1
2.1
2.5
80
2.3
2.3
2.9
90
2.5
2.5
3.2
100
2.7
2.7
3.6
110
3.0
3.0
3.9
120
3.2
3.2
4.3
130
3.5
3.5
4.7
140
3.8
3.8
5.0
150
4.1
4.1
5.4
(2)
For capacitor energy storage equipment, compliance with the requirements of this
subsection shall be determined with the system fully charged and a setting of
ten mAs for each exposure.
(3)
The required minimal aluminum equivalent filtration shall include the filtration
contributed by all materials which are permanently present between the source and
the patient.
(4)
For mammography systems with molybdenum filter and molybdenum target,
measured half-value layer (HVL) with compression device in the x-ray beam shall
be greater than or equal to the kilovolts peak (kVp) divided by one hundred,
millimeters aluminum and less than or equal to the kilovolts peak (kVp) divided by
one hundred plus one-tenth millimeter aluminum.
HVL
> (kVp/100) mmAl and
< (kVp/100) + 0.1 mmAl
b.
Filtration controls. For x-ray systems which have variable kilovolts peak and variable
filtration for the useful beam, a device shall link the kilovolts peak selector with the filters
and shall prevent an exposure unless the minimum amount of filtration required by
paragraph 1 of subdivision a is in the useful beam for the given kilovolts peak which has
been selected.
6.
Multiple tubes. Where two or more radiographic tubes are controlled by one exposure switch,
the tube or tubes which have been selected shall be clearly indicated prior to initiation of the
exposure. This indication shall be both on the x-ray control panel and at or near the tube
housing assembly which has been selected.
7.
Mechanical support of tube head. The tube housing assembly supports shall be adjusted
such that the tube housing assembly will remain stable during an exposure unless tube
housing movement is a designed function of the x-ray system.
8.
Technique indicators.
a.
The technique factors to be used during an exposure shall be indicated before the
exposure begins, except when automatic exposure controls are used, in which case the
technique factors which are set prior to the exposure shall be indicated.
b.
The requirements of subdivision a may be met by permanent markings on equipment
having fixed technique factors. Indication of technique factors shall be visible from the
operator's position except in the case of spot films made by the fluoroscopist.
9.
Maintaining compliance. Diagnostic x-ray systems and their associated components used
on humans and certified pursuant to the federal x-ray equipment performance standard
(21 CFR part 1020) shall be maintained in compliance with applicable requirements of that
standard.
10.
Locks. All position locking, holding, and centering devices on x-ray system components and
systems shall function as intended.
11.
Structural shielding requirements (see appendix C).