NDAC 33.1-10-06-03
General requirements
Cite as N.D. Admin. Code ยง 33.1-10-06-03
1.
Administrative controls.
a.
Registrant. The registrant shall be responsible for directing the operation of the x-ray
systems which have been registered with the department. The registrant or the
registrant's agent shall assure that the requirements are met in the operation of the x-ray
system.
(1)
An x-ray system which does not meet the requirements of this article shall not be
operated for diagnostic or therapeutic purposes.
(2)
All individuals, except those listed in part 1 of appendix G, prior to operating the
x-ray systems shall be adequately instructed in the safe operating procedures and
be competent in the safe use of the equipment commensurate with the size, scope,
and nature of the service as outlined in appendix F. In addition, all individuals shall
meet the specific requirements as outlined in subparagraph a or b. The department
may use interview, observation, or testing to determine compliance. Records must
be maintained by the registrant to demonstrate compliance with this paragraph.
(a)
General diagnostic operators are not limited in scope of practice. Obtaining
general diagnostic operator status will consist of one of the following:
[1]
Obtain board eligibility or board certification with the American registry of
radiologic technologists (ARRT);
[2]
Obtain board eligibility or board certification with the American chiropractic
registry of radiologic technologists (ACRRT); and only perform x-ray
examinations for chiropractic services;
[3]
Receive department approval, through individual consideration, by
demonstration of an acceptable level of education and clinical training; or
[4]
Demonstrate current enrollment in an educational program accredited by
a process acceptable to the department, and provide documentation of
competency in all routine radiographic procedures and specialty views.
(b)
Limited x-ray machine operators are limited in scope of practice to only those
procedures listed in appendix I, except as allowed in subparagraph c. Limited
x-ray machine operators must meet the prerequisite qualifications, receive
training, and demonstrate competence as follows:
[1]
Limited x-ray machine operators shall have successfully completed the
course of training required by one of the allied health professions listed in
part 2 of appendix G;
[2]
Complete at least eighty hours of didactic instruction at a single training
program providing didactic instruction in accordance with part 1 of
appendix H;
[3]
Complete the three hour self-study course designed by the department of
environmental quality; and
[4]
Complete the clinical experience requirements in part 2 of appendix H.
(c)
Limited x-ray machine operators may only conduct diagnostic x-ray
examinations outside the scope of practice of appendix I in accordance with
the following:
[1]
When it is determined to be an emergency and ordered by individuals
listed in part 3 of appendix G. The individual requesting the procedures
must comply with subitems a, b, and c.
[a]
The requesting individual must provide a written order specifying
what types of diagnostic x-ray examinations outside the scope of
procedures listed in appendix I are requested. The order shall
contain an explanation of the emergency nature or medical reason
for the order.
[b]
The requesting individual must provide direct supervision during the
time the x-ray image is obtained.
[c]
The facility must keep records of all emergency x-ray procedures
ordered under this subparagraph.
[2]
When a practice requires a specific view or examination outside the
scope of practice listed in appendix I to be conducted on a routine basis,
and the facility has only limited x-ray machine operators, application may
be made to the department requesting approval for a limited x-ray
machine operator to perform the procedure. This allowance shall be
limited to the facility, the specific individual, and the procedure requested.
After an allowance has been granted, reapplication and reauthorization
are not necessary for the same procedure. The application for allowance
should include the following:
[a]
Documentation which demonstrates the need for the specific view;
[b]
Documentation on forms supplied by the department indicating that
each individual for which the request is made has demonstrated
competence in the procedure; and
[c]
Proof of additional didactic instruction or completion of examination
as deemed necessary by the department for each individual.
[3]
When it is not a computed tomography examination.
(d)
Limited x-ray machine operator implementation period. Individuals who begin taking
x-rays after one year from March 1, 2003, will have to meet all of the requirements
of this paragraph before operating the x-ray system.
(3)
General diagnostic and limited diagnostic x-ray operators shall maintain continuing
education units as outlined in appendix K.
(4)
A chart shall be provided in the vicinity of the diagnostic x-ray system's control
panel, which specifies for all examinations performed with that system the following
information:
(a)
Patient's body part and anatomical size or body thickness, or age (for
pediatrics), versus technique factors to be utilized.
(b)
Type and size of the film or film screen combination to be used.
(c)
Type and focal distance of the grid to be used, if any.
(d)
Source-image receptor distance to be used (except for dental intraoral
radiography).
(e)
Type and location of placement of gonad shielding to be used.
(f)
For mammography, indication of kVp/target/filter combination.
(5)
The registrant of a facility shall create and make available to x-ray operators written
safety procedures, including patient holding restrictions and any restrictions of the
operating technique required for the safe operation of the particular x-ray system.
The operator shall be able to demonstrate familiarity with these procedures.
(6)
Except for human patients who cannot be moved out of the room, only the staff and
ancillary personnel or other persons required for the medical procedure or training
shall be in the room during the radiographic exposure. Other than the patient being
examined:
(a)
All individuals shall be positioned such that no part of the body will be struck by
the useful beam unless protected by not less than five-tenths millimeter lead
equivalent material.
(b)
The x-ray operator, other staff, ancillary personnel, and other persons required
for the x-ray procedure shall be protected from the direct scatter radiation by
protective aprons or whole body protective barriers of not less than twenty-five
one-hundredths millimeter lead equivalent material.
(c)
Human patients who cannot be removed from the room shall be protected from
the direct scatter radiation by whole body protective barriers of not less than
twenty-five one-hundredths millimeter lead equivalent material or shall be so
positioned that the nearest portion of the body is at least two meters from both
the tube head and the nearest edge of the image receptor.
(7)
Gonad shielding of not less than five-tenths millimeter lead equivalent material must
be used for human patients who have not passed the reproductive age during
radiographic procedures in which the gonads are in the useful beam, except for
cases in which this would interfere with the diagnostic procedure.
(8)
Individuals may not be exposed to the useful beam except for healing arts purposes
and when such exposure has been authorized by a licensed practitioner of the
healing arts. This provision specifically prohibits deliberate exposure for the
following purposes:
(a)
Exposure of an individual for training, demonstration, or other non-healing-arts
purposes.
(b)
Exposure of an individual for the purpose of healing arts screening except as
authorized by paragraph 12.
(9)
When a patient or film must be provided with auxiliary support during a radiation
exposure:
(a)
Mechanical holding devices shall be used when the technique permits. The
safety rules, required by this section shall list individual projections where
holding devices cannot be utilized.
(b)
Written safety procedures, as required by paragraph 4, shall indicate the
requirements for selecting a holder and the procedure the holder shall follow.
(c)
The human holder shall be instructed in personal radiation safety and
protected as required by paragraph 6.
(d)
No individual shall be used routinely to hold film or patients.
(e)
In those cases where the patient must hold the film, except during intraoral
examinations, any portion of the body other than the area of clinical interest
struck by the useful beam shall be protected by not less than five-tenths
millimeter lead equivalent material.
(f)
A record shall be made of the examination and shall include the name of the
human holder, date of the examination, number of exposures, and technique
factors utilized for the exposure.
(g)
Each facility shall have leaded aprons and gloves available in sufficient
numbers to provide protection to all personnel who are involved with x-ray
operations and who are otherwise not shielded.
(10)
Procedures and auxiliary equipment designed to minimize patient and personnel
exposure commensurate with the needed diagnostic information shall be utilized.
This is interpreted to include:
(a)
The speed of film and screen combinations shall be the fastest speed
consistent with the diagnostic objective of the examinations. Film cassettes
without intensifying screens shall not be used for any routine diagnostic
radiological imaging, with the exception of veterinary radiography, therapeutic
portal imaging, and standard film packets for intraoral use in dental
radiography.
(b)
The radiation exposure to the patient shall be the minimum exposure required
to produce images of good diagnostic quality.
(c)
Proper film handling and processing procedures. Each installation using a
radiographic x-ray system and using analog image receptors (e.g.,
radiographic film) shall have available suitable equipment for handling and
processing radiographic film in accordance with appendix D.
(d)
Portable or mobile equipment shall be used only for examinations where it is
impractical to transfer the patients to a stationary x-ray installation.
(e)
X-ray systems subject to section 33.1-10-06-06 shall not be utilized in
procedures where the source to patient distance is less than thirty centimeters,
except for veterinary systems.
(f)
If grids are used between the patient and the image receptor to decrease
scatter to the film and improve contrast, the grid shall:
[1]
Be positioned properly, for example, tube side facing the right direction
and grid centered to the central ray; and
[2]
If the grid is of the focused type, be of the proper focal distance for the
source-image distances being used.
(11)
All individuals who are associated with the operation of an x-ray system are subject
to the requirements of section 33.1-10-04.2-01 [10 CFR 20.1201, 20.1207, and
20.1208]. In addition:
(a)
When protective clothing or devices are worn on portions of the body and a
monitoring device is required, at least one such monitoring device shall be
utilized as follows:
[1]
When an apron is worn, the monitoring device shall be worn at the collar
outside of the apron.
[2]
The dose to the whole body based on the maximum dose attributed to the
most critical organ shall be recorded in the reports required by section
33.1-10-04.2-01 [10 CFR 20.2206]. If more than one device is used and a
record is made of the data, each dose shall be identified with the area
where the device was worn on the body.
(b)
Exposure of a personnel monitoring device to deceptively indicate a dose
delivered to an individual is prohibited.
(12)
Healing arts screening. Any person proposing to conduct a healing arts screening
program shall not initiate such a program without prior approval of the department.
When requesting such approval, that person shall submit the information outlined in
appendix E. If any information submitted to the department becomes invalid or
outdated, the department shall be immediately notified.
b.
Information and maintenance record and associated information. The registrant shall
maintain the following information for each x-ray system for inspection by the
department:
(1)
Maximum rating of technique factors.
(2)
Model and serial numbers of all major components and user's manuals for those
components.
(3)
Aluminum equivalent filtration of the useful beam, including any routine variation.
(4)
Tube rating charts and cooling curves.
(5)
Records of surveys, calibrations, maintenance, and modifications performed on the
x-ray system with the names of persons who performed such services.
(6)
A scale drawing of the room in which a stationary x-ray system is located with such
drawing indicating the use of areas adjacent to the room and an estimation of the
extent of occupancy by an individual in such areas. In addition, the drawing shall
include:
(a)
The results of a survey for radiation levels present at the operator's position
and at pertinent points outside the room at specified test conditions; or
(b)
The type and thickness of materials, or lead equivalency, of each protective
barrier.
(7)
A copy of all correspondence with this department regarding that x-ray system.
c.
X-ray log.
(1)
Except for veterinary facilities, each facility shall maintain an x-ray log containing the
patient's name, the type of examinations, the dates those examinations were
performed, and the name of the x-ray operator. When the patient or film must be
provided with human auxiliary support, the name of the human holder shall be
recorded.
(2)
Veterinary facilities shall maintain an x-ray utilization log indicating the type of
examinations, the date of the examinations and if the patient or film was provided
with human auxiliary support, the name of the human holder.
2.
Plan review.
a.
Prior to construction, the floor plans, shielding specifications, and equipment
arrangement of all new installations, or modifications of existing installations, utilizing
ionizing radiation machines shall be submitted to the department for review and approval.
The required information is denoted in appendices A, B, and C.
b.
The department may require the applicant to utilize the services of a qualified expert to
determine the shielding requirements prior to the plan review and approval.
c.
The approval of such plans shall not preclude the requirement of additional modifications
should a subsequent analysis of operating conditions indicate the possibility of an
individual receiving a dose in excess of the limits prescribed in section 33.1-10-04.2-01
[10 CFR 20.1201, 20.1207, 20.1208, and 20.1301].