NDAC 33.1-10-08-04
Area requirements
Cite as N.D. Admin. Code ยง 33.1-10-08-04
1.
Radiation levels. The local components of an analytical x-ray system shall be located and
arranged and shall include sufficient shielding or access control such that no radiation levels
exist in any area surrounding the local component group which could result in a dose to an
individual present therein in excess of the dose limits given in chapter 33.1-10-04.2. For
systems utilizing x-ray tubes, these levels shall be met at any specified tube rating.
2.
Surveys.
a.
Radiation surveys, as required by chapter 33.1-10-04.2, of all analytical x-ray systems
sufficient to show compliance with subsection 1 shall be performed:
(1)
Upon installation of the equipment, and at least once every twelve months
thereafter.
(2)
Following any change in the initial arrangement, number, or type of local
components in the system.
(3)
Following any maintenance requiring the disassembly or removal of a local
component in the system.
(4)
During the performance of maintenance and alignment procedures if the procedures
require the presence of a primary x-ray beam when any local component in the
system is disassembled or removed.
(5)
Any time a visual inspection of the local components in the system reveals an
abnormal condition.
(6)
Whenever personnel monitoring devices show a significant increase over the
previous monitoring period or the readings are approaching the limits specified in
chapter 33.1-10-04.2.
b.
Radiation survey measurements shall not be required if a registrant can demonstrate
compliance with subsection 1 to the satisfaction of the department.
3.
Posting. Each area or room containing analytical x-ray equipment shall be conspicuously
posted with a sign or signs bearing the radiation symbol and the words "CAUTION - X-RAY
EQUIPMENT", or words having a similar intent in accordance with chapter 33.1-10-04.2.