NDAC 33.1-10-08-03
Equipment requirements
Cite as N.D. Admin. Code ยง 33.1-10-08-03
1.
Safety device. A device which prevents the entry of any portion of an individual's body into
the primary x-ray beam path or which causes the beam to be shut off upon entry into its path
shall be provided on all open-beam configurations. A registrant may apply to the department
for an exemption from the requirement of a safety device. Such application shall include:
a.
A description of the various safety devices that have been evaluated.
b.
The reason each of these devices cannot be used.
c.
A description of the alternative methods that will be employed to minimize the possibility
of an accidental exposure, including procedures to assure that operators and others in
the area will be informed of the absence of safety devices.
2.
Warning devices.
a.
Open-beam configurations shall be provided with a readily discernible indication of:
(1)
X-ray tube (ON-OFF) status located near the radiation source housing, if the
primary beam is controlled in this manner.
(2)
Shutter status (OPEN-CLOSED) located near each port on the radiation source
housing, if the primary beam is controlled in this manner.
b.
An easily visible warning light labeled with the words "X-RAY ON", or word sharing a
similar intent, must be located:
(1)
Near any switch that energizes an x-ray tube and shall be illuminated only when the
tube is energized.
(2)
In the case of a radioactive source, near any switch that opens a housing shutter
and must be illuminated only when the shutter is open.
c.
Warning devices shall be labeled so that their purpose is easily identified. On equipment
installed after August 1, 1979, warning devices shall have fail-safe characteristics.
3.
Ports. Unused ports on radiation source housings shall be secured in the closed position in a
manner which will prevent casual opening.
4.
Labeling. All analytical x-ray equipment shall be labeled with a readily discernible sign or
signs bearing the radiation symbol and the words:
a.
"CAUTION - HIGH-INTENSITY X-RAY BEAM", or words having a similar intent, on the
x-ray source housing; and
b.
"CAUTION RADIATION - THIS EQUIPMENT PRODUCES RADIATION WHEN
ENERGIZED", or words having a similar intent, near any switch that energizes an x-ray
tube if the radiation source is an x-ray tube; or
c.
"CAUTION - RADIOACTIVE MATERIAL", or words having a similar intent, on the source
housing in accordance with chapter 33.1-10-04.2 if the radiation source is a radionuclide.
5.
Shutters. On open-beam configurations installed after August 1, 1979, each port on the
radiation source housing shall be equipped with a shutter that cannot be opened unless a
collimator or a coupling has been connected to the port.
6.
Warning lights.
a.
An easily visible warning light labeled with the words "X-RAY ON", or words having a
similar intent, shall be located:
(1)
Near any switch that energizes an x-ray tube and shall be illuminated only when the
tube is energized; or
(2)
In the case of a radioactive source, near any switch that opens a housing shutter,
and shall be illuminated only when the shutter is open.
b.
On equipment installed after August 1, 1979, warning lights shall have fail-safe
characteristics.
7.
Radiation source housing. Each radiation source housing is subject to the following
requirements:
a.
Each x-ray tube housing shall be equipped with an interlock that shuts off the tube if it is
removed from the radiation source housing or if the housing is disassembled.
b.
Each radioactive source housing or port cover or each x-ray tube housing shall be so
constructed that, with all shutters closed, the radiation measured at a distance of five
centimeters from its surface is not capable of producing a dose in excess of twenty-five
hundredths millisieverts [2.5 millirems] in one hour. For systems utilizing x-ray tubes, this
limit shall be met at any specified tube rating.
8.
Generator cabinet. Each x-ray generator shall be supplied with a protective cabinet which
limits leakage radiation measured at a distance of five centimeters from its surface such that it
is not capable of producing a dose in excess of two and one-half microsieverts [0.25 millirem]
in one hour.