OAC 310:681-8-4
Laboratory quality assurance and quality control
Cite as Okla. Admin. Code § 310:681-8-4
(a) Laboratory Quality Assurance (LQA) program. The medical
laboratory director shall develop and implement an LQA program to
ensure the reliability and validity of the analytical data produced by
the laboratory.
(1) The LQA program shall, at minimum, include a written LQA manual
that addresses the following:
(A) Quality control procedures, including remedial actions;
(B) Laboratory organization and employee training and
responsibilities;
(C) LQA criteria for acceptable performance;
(D) Traceability of data and analytical results;
(E) Instrument maintenance, calibration procedures, and
frequency;
(F) Performance and system audits;
(G) Steps to change processes when necessary;
(H) Record retention;
(I) Test procedure standardization; and
(J) Method validation.
(2) The laboratory director shall annually review, amend if
necessary, and approve the LQA program and manual when:
2021.11.23_OSDH681_OMMARulesChapterAsAmended
99
(A) The LQA program and manual are created; and
(B) There is a change in methods, laboratory equipment, or the
supervisory or management laboratory employee overseeing the LQA
program.
(b) Laboratory quality control samples.
(1) The laboratory shall use laboratory quality control (LQC)
samples in the performance of each analysis according to the
specifications in this section.
(2) The laboratory shall analyze LQC samples in the same manner as
the laboratory analyzes samples of medical marijuana and medical
marijuana products.
(3) The laboratory shall use negative and positive controls for
microbial testing.
(4) The following quality control samples must be run every 20
samples in an analytic run:
(A) Method blank;
(B) Continuing calibration verification (CCV);
(C) Laboratory replicate sample; and
(D) Matrix spike sample or matrix spike duplicate sample.
(5) If the result of the analyses is outside the specified
acceptance criteria in Appendix B, the laboratory shall determine
the cause and take steps to remedy the problem until the result is
within the specified acceptance criteria. Samples after the last
acceptable run must be re-tested.
(6) The laboratory shall generate a LQC sample report for each
analytical run that includes LQC parameters, measurements, analysis
date, and matrix. The results must fall within the criteria set
forth in Appendix B.
(c) Reagents, solutions, and reference standards.
(1) Reagents, solutions, and reference standards shall be:
(A) Secured in accordance with the laboratory's storage
policies; labeled to indicate identity, date received or
prepared, and expiration or requalification date; and, where
applicable, concentration or purity, storage requirements, and
date opened;
(B) Stored under appropriate conditions to minimize
degradation or deterioration of the material; and
(C) Used only within the item's expiration or requalification
date.
(2) Deteriorated or outdated reagents and solutions shall be
properly disposed, in compliance with all federal, state and local
regulations.
(3) The laboratory may acquire commercial reference standards for
cannabinoids and other chemicals or contaminants, for the exclusive
purpose of conducting testing for which the laboratory is approved.
The laboratory may elect to produce reference standards in-house
(internally). When internally produced, the laboratory shall utilize
standard analytical techniques to document the purity and
concentration of the internally produced reference standards. The
laboratory is authorized to obtain marijuana or marijuana-derived
product from a licensed non-profit producer for this purpose.
(4) The laboratory shall obtain or, for internally-produced
2021.11.23_OSDH681_OMMARulesChapterAsAmended
100
standards, shall create a certificate of analysis (COA) for each lot
of reference standard. Each COA shall be kept on-file and the lot
number of the reference standard used shall be recorded in the
documentation for each analysis, as applicable.