WAC 246-240-617. Records of periodic spot-checks for teletherapy units

Last amended: 2006Year: 2026Length: 192 wordsOfficial source
(1) A licensee shall retain a record of each periodic spot-check for teletherapy units required by WAC 246-240-378 for three years. (2) The record must include: (a) The date of the spot-check; (b) The manufacturer's name, model number, and serial number of the teletherapy unit, source and instrument used to measure the output of the teletherapy unit; (c) An assessment of timer linearity and constancy; (d) The calculated on-off error; (e) A determination of the coincidence of the radiation field and the field indicated by the light beam localizing device; (f) The determined accuracy of each distance measuring and localization device; (g) The difference between the anticipated output and the measured output; (h) Notations indicating the operability of each entrance door electrical interlock, each electrical or mechanical stop, each source exposure indicator light, and the viewing and intercom system and doors; and (i) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check. (3) A licensee shall retain a copy of the procedures required by WAC 246-240-378 (2) until the licensee no longer possesses the teletherapy unit.