(1) A licensee shall retain a record of each spot-check for remote afterloader units required by WAC 246-240-381 for three years. (2) The record must include, as applicable: (a) The date of the spot-check; (b) The manufacturer's name, model number, and serial number for the remote afterloader unit and source; (c) An assessment of timer accuracy; (d) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and (e) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check. (3) A licensee shall retain a copy of the procedures required by WAC 246-240-381 (2) until the licensee no longer possesses the remote afterloader unit.